31
8. Error Message
The error message will be displayed at the right of the
time display as shown left.
The initial letter of each error will be displayed according
to the priority order of errors. (Priority order: L
→
F
→
O)
Error message
Content of error
Action
L Laser
Error
Failure of the laser luminescence part.
Please contact your local distributor or our service center for
information.
F Flow
Error
Displayed when the absorption flow rate is out of the specified range
of 2.83L/min
±
10%.
If a filter or a tube is attached to the inlet of the instrument, please
remove it. If the “F” error sign still remains, it is a failure of the flow
route system including the pump.
Please contact your local distributor or our service center for
information. (Please refer to section 3.3.2.)
O
Maximum Concentration
Exceeded
Displayed when the measurable concentration of the instrument is
exceed.
Please move to a cleaner place, or install the filter and measure.
If the “O” error sign still remains, please contact your local distributor
or our service center for information.
M
Printer Buffer Exceeded
Displayed when the printer buffer is exceeded
Please note that once this error is displayed, the remaining data will
not be displayed.
<<Caution>>
When the measurement interval (INT) in REPEAT mode is set at 15 seconds or less, there is a
possibility of Buffer error when printing is executed during a measurement.
To print during a measurement, please set the measurement interval over 15 seconds.
REPEAT 003BPM 15:25 L
WAIT 0.3 0.00E+0
/
cf
01
/
06 0.5 0.00E+0
/
cf
25.5
℃
56.0% 0.25m
/
s
Error display location (L, F, O)
Error display location (M)