15
⑫
LARGE particle size
⑩
SMALL particle size
5. Measurement Method
5.1 Explanation of measurement screen
Name
Explanation
1 Measuring
mode
Five kinds of REPEAT, SINGLE, CONT, CALC,
REMOTE
2
Number of data records
003: Storage No. No display : No data records
(e.g. 003: three data records)
3
Warning beep sound
B: Beep sounded
Display no: Beep not sounded
4
Data printing
P: Data printed
Display no: Data not printed (refer to 6.4)
5 Error
message
M: The buffer memory is over loaded when printing the data
(refer to 8)
6
Current time
Refer to 4.5
7 Error
message
F: Flow error
L: Laser error
O
:
Maximum concentration is exceeded (refer to 8)
8
Status of measurement
WAIT: Starting up
READY: Ready for measurement
STOP: Measurement finished (refer to 5.2-5.6)
9 Sampling
time
Tag number of the current measurement/ the specified sampling
frequency.
10 SMALL particle size
Smaller of 2 particle sizes chose at setup screen.
11 SMALL
data
The number of counts or concentration of the particle size on 10.
(refer to 4.4)
12 LARGE particle size
Lager of 2 particle sizes chose at setup screen.
13 LARGE
data
The number of counts or concentration of the particle size on 10.
(refer to 4.4)
14 Temperature
data
Show the data by selected the T/H probe uses (refer to 4.4
)
15 Humidity
data
Show the data by selected the T/H probe uses (refer to 4.4
)
16 Air velocity data
Show the data by selected the Air velocity probe uses (refer to
4.4
)
③
Warning beep sound
①
Measuring mode
REPEAT 003BPM 15:25
□
WAIT 0.3 0.00E+0
/
cf
01
/
06 0.5 0.00E+0
/
cf
25.5
℃
56.0% 0.25m
/
s
⑦
Error message
⑧
Status of measurement
②
Number of data records
⑪
SMALL data
⑨
Sampling number
⑬
LARGE data
⑭
Temperature data
⑯
Air velocity data
④
Data printing
⑤
Error message
⑥
Current time
⑮
Humidity data