25
6.4 Printout of stored data… <4.DATA>
→
<3.PRINT>
Put the optional Printer cable into the modular jack of GEO--
α
,
and connect other side of the cable with the Printer to print
the stored data.
1
START
The first tag number of the stored data
to be printed
2
END
The last tag number of the stored data
to be printed
△
▽
key
Adjust the value
SET key
Move the cursor
START/STOP key
Start printing of the stored data
PREV
key
Go back to DATA
Preparations
Printer (Option)
・・・・・・・・
Recommendatory Printer Model. DPU-201GS (SEIKO CO., LTD),
Signal cable (Model 3886-07: Option), Signal cable connect GEO-
α
with Printer
DIP Switch Setting
Switch Number
Function
GEO-
α
Setting
Printer
SW1
Word length
8 bit
ON
SW2
Parity bit
None
ON
SW3
Set parity
Odd number
ON
SW4
〜
6
Baud rate
9600
following table
Baud rate
SW4 Setting
SW5 Setting
SW6 Setting
9600 OFF OFF ON
* When using DPU-H245, please use it with the manufacturers default setting.
Signal cable
GEO-
α
Printer
Pin number
Signal name
Pin number
Signal name
1 TXD 3 DATA
6 GND 4 GND
5 CTS 8 BUSY
6 GND 5 GND
<<Caution>>
When the measurement interval (INT) in REPEAT mode is set at 15 seconds or less, there is a possibility
of Buffer error when printing is executed during a measurement. To print during a measurement, please
set the measurement interval over 15 seconds.
Modular jack
Start printing
<PRINT>
STORE 139
1. START 001
2. END 139