16
5.2 REPEAT Mode
By setting the sampling time, frequency and interval of each measurement, this unit automatically measures as
specified and stops after measurements. Interval is the time between the beginning of first measurement and the
next. The setting of the particle size (um), data storage (STR), alarm (BEEP) and printout (PR) are possible.
Display screen
Operation
key
Operation explanation
POWER
Push POWER key to input the power supply.
PREV
Push PREV key twice to proceed MODE screen
△
▽
SET Select 1.REPEAT
△
▽
SET
■
Setup the particle size, requirement of data
storage , alarm , printout, sampling time,
frequency and interval of measurement.
Use
△
▽
key to change the setting condition,
then push SET key.
After the input is done, change NO to OK
and push SET key.
△
▽
SET
①:
sum total of sampling time
②:
total length of time from the beginning to
the end of measurement
After confirming these TIMEs, change NO to OK and
push SET key.
WAIT Mode
WAIT sign is shown for the stabilization of internal
pump.
READY Mode
WAIT sign turns to READY in 10 seconds and
measurement can be started.
START
/
STOP
Push START/STOP key to start the measurement.
The display shows the real-time data.
③:
remaining time of each measurements
④:
measurement number
Interval menu
Screen changes to interval mode after the sampling
time are over.
⑤:
starting time of next measurement
Measuring
Measurement is automatically started from the
indicated starting time.
★
The data is printed after the sampling time is over if
you chose printout. (refer to 6.4)
★
To stop the measurement halfway
Push START/STOP key.
Only the data of finished previous
measurement before pushing the STOP key
will be stored if you chose data storage
MEASURING
SAMPLE TIME 5:00:00
TOTAL TIME 14:40:00
OK
REPEAT 003BPM 15:25
WAIT 0.3 0.00E+0
/
cf
01
/
30 0.5 0.00E+0
/
cf
REPEAT 003BPM 15:25
READY 0.3 0.00E+0
/
cf
01
/
30 0.5 0.00E+0
/
cf
NEXT 004BPM 15:35
15:55 0.3 0.00E+0
/
cf
02
/
30 0.5 0.00E+0
/
cf
REPEAT
0.3
/
0.5µm
STR:Y BEEP:N PR:N
SAMPLE 10:00 30TIMES
INT 00:30:00 OK
<MODE>
1. REPEAT 4.CALC
2. SINGLE 5.REMOTE
3. CONT
←
①
←
②
⑤
→
REPEAT 004BPM 15:25
09:59 0.3 0.00E+0
/
cf
01
/
30 0.5 0.00E+0
/
cf
③
→
④
→
REPEAT 004BPM 15:55
09:59 0.3 0.00E+0
/
cf
02
/
30 0.5 0.00E+0
/
cf