24
(2) Calculation mode
Format Byte
Explanation
999 crlf
5
Store No
9 crlf 3
Measurement mode (4
:
Calculation
)
99,99,99 crlf 10
Measurement start date
99,99,99 crlf 10
Measurement start time
99999 crlf 7
Measurement number
99,99,99 crlf 10
Sampling time
(
hours, minutes, seconds
)
xxx crlf 5
Particle unit
(
CNT , /cf , /m
3
)
x crlf 4
Temperature unit
(
C , F
)
xxx crlf 5
Air velocity unit
(
m/s , FPM
)
x,x,x crlf
7
Error message
(L
:
Light source, F
:
Flow rate, O
:
Over the maximum concentration)
9.999E+99, 10
Average of 0.3µm
9.999E+99, 10
Standard deviation of 0.3µm
999999999, 10
Maximum data of 0.3µm, 9.999E+99 crlf using /cf or /m
3
as unit
999999999 crlf
11
Minimum data of 0.3µm, 9.999E+99 crlf using /cf or /m
3
as unit
9.999E+99, 10
Average of 0.5m
9.999E+99, 10
Standard deviation of 0.5µm
999999999, 10
Maximum data of 0.5µm, 9.999E+99 crlf using /cf or /m
3
as unit
999999999 crlf
11
Minimum data of 0.5µm, 9.999E+99 crlf using /cf or /m
3
as unit
9.999E+99, 10
Average of 1µm
9.999E+99, 10
Standard deviation of 1µm
999999999, 10
Maximum data of 1µm, 9.999E+99 crlf using /cf or /m
3
as unit
999999999 crlf
11
Minimum data of 1µm, 9.999E+99 crlf using /cf or /m
3
as unit
9.999E+99, 10
Average of 3µm
9.999E+99, 10
Standard deviation of 3µm
999999999, 10
Maximum data of 3µm, 9.999E+99 crlf using /cf or /m
3
as unit
999999999 crlf
11
Minimum data of 3µm, 9.999E+99 crlf using /cf or /m
3
as unit
9.999E+99, 10
Average of 5µm
9.999E+99, 10
Standard deviation of 5µm
999999999, 10
Maximum data of 5µm, 9.999E+99 crlf using /cf or /m
3
as unit
999999999 crlf
11
Minimum data of 5µm, 9.999E+99 crlf using /cf or /m
3
as unit
999.9, 6
Average of Temperature
999.9, 6
Temperature data standard deviation
999.9, 6
Maximum data of Temperature
999.9 crlf
7
Minimum data of Temperature
999.9, 6
Average of Humidity
999.9, 6
Standard deviation of Humidity
999.9, 6
Maximum data of Humidity
999.9 crlf
7
Minimum data of Humidity data
9.999, 6
Average of Air velocity 999.9 using FRM as unit
9.999, 6
Standard deviation of Air velocity 999.9 using FRM as unit
9.999, 6
Maximum data of Air velocity 999.9 using FRM as unit
9.999 crlf
7
Minimum data of Air velocity 999.9 using FRM as unit
Total
346
*)
・
T./H or Air velocity probe is not selected it becomes “ ***** crlf”.
・
It becomes “###.#” when the value of T/H probe exceeds measurement range.
・
It becomes “###.#” using m/s as unit
(when the value of Air velocity probe exceeds measurement range.)