3. USING THE FOUR FUNCTION BUTTONS:
Switching ON
the Analyser
Rotate the dial to the mode you want to use before switching on.
This may eliminate the need for a full countdown in some of the
modes and save you time.
Press
/
button to switch the unit ON. This must be
done in fresh air to ensure that the analyser auto calibrates its
sensors properly.
When switched on, the analyser beeps and briefly displays
software version, date and time. Its bottom line counts down until
the sensors are ready to use. If the analyser will not auto calibrate,
its sensors need to be replaced or recalibrated by an authorised
repair centre.
If an inlet temperature probe (optional) is connected into the T2
socket during its countdown, the measured temperature from the
inlet probe will be used as the inlet temperature.
If an inlet temperature probe is not connected to the analyser
during countdown the measured temperature from the flue probe
will be used as the inlet temperature.
If neither probe is connected during countdown the analyser’s
internal ambient temperature will be used as the inlet temperature.
Switching OFF
the Analyser
Press
/
& hold for 2 seconds to switch the analyser
OFF. The display counts down from 30 or less with the pump on to
clear the sensors with fresh air – If the probe is still connected,
make sure analyser and probe are in fresh air.
Press
/
if you want to stop the countdown and return
to making measurements.
Note: The analyser will not switch off unless the CO
reading is below 40ppm.
Torch Light
Press
/
to switch the torch light on and off.
NOTE: Use of the torch light significantly increases the
current drain on the batteries.