background image

 

 

Page 24

 

 

 

SWEEP TEST 

 

TYPE C SEALED 
APPLIANCE 

 

TIGHTNESS TEST 

 

 

 

 

     

 

 

 

 

 

 

KANE456      SW19392 

V0.12    KANE456     SW19392 

V0.12   

KANE456      SW19392 

V0.12 

 

 

 

   

 

 

 

 

 

 

YOUR COMPANY NAME & 

  YOUR COMPANY NAME & 

 

YOUR COMPANY NAME & 

PHONE NUMBER HERE 

  PHONE NUMBER HERE 

 

PHONE NUMBER HERE 

 

 

 

   

 

 

 

 

 

 

SERIAL NO.              9876543210    SERIAL NO.              9876543210   

SERIAL NO.              9876543210 

 

 

 

   

 

 

 

 

 

 

LOG NO.   

06    LOG NO.   

03   

DATE 

 

19/01/16 

 

 

 

   

 

 

 

TIME 

 

11:09:16 

DATE 

 

19/01/16 

  DATE 

 

19/01/16 

 

 

 

 

TIME 

 

10:11:11 

  TIME 

 

12:25:27 

 

CAL DUE ON                  18/12/16 

 

 

 

   

 

 

 

 

 

 

CAL DUE ON                  18/12/16    CAL DUE ON                  18/12/16   

LET BY TEST 

 

 

 

   

 

 

 

PRS 1 

mbar 

10.80 

ROOM CO 

  ROOM CO 

 

PRS 2 

mbar 

10.78 

SWEEP TEST 

  TYPE C 

 

 

 

LET BY 

MINS 

1:00 

 

 

 

  SEALED 

 

 

 

 

 

 

LIMIT 

 

10ppm    APPLIANCE 

   

TIGHTNESS TEST 

ALARM 

 

30ppm     

 

   

PRS 1 

mbar 

20.14 

TESTS 

 

1    LIMIT 

 

10ppm   

PRS 2 

mbar  

20.13  

 

 

    ALARM 

 

30ppm   

mbar  

-0.01  

 

CO ppm    TESTS 

 

15   

 

01 

 

0     

 

 

 

STABILIS’N 

MINS 

1:00 

MAXIMUM CO ppm                    0    TEST 

 

CO ppm   

TIGHTNESS 

MINS 

2:00 

 

 

    01 

 

0   

 

 

 

    02 

 

0   

CUSTOMER 

CUSTOMER 

  03 

 

0   

 

 

    04 

 

0   

 

 

 

    05 

 

0   

 

 

 

    06 

 

0   

APPLIANCE 

APPLIANCE 

  07 

 

0   

 

 

 

    08 

 

0   

 

 

 

    09 

 

0   

 

 

 

 

 

   

10 

 

 

REFERENCE 

REFERENCE 

  11 

 

0   

 

 

  12 

 

   

 

 

 

 

 

  13 

 

0   

 

 

 

 

 

 

  14 

 

0   

 

 

 

 

 

 

  15 

 

0   

 

 

 

 

 

 

  MAXIMUM CO ppm 

0   

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

  CUSTOMER 

 

 

 

 

 

   

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

  APPLIANCE 

 

 

 

 

 

   

 

 

 

 

 

   

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

  REFERENCE 

 

 

 

 

 

   

 

 

 

 

 

   

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

Содержание 456

Страница 1: ...KANE456 Flue Gas Analyser with direct CO2 measurement Stock No 19738 2 February 2017 Kane International Ltd...

Страница 2: ...OUR FUNCTION BUTTONS 10 11 4 USING THE ANALYSER 12 21 4 1 COMBUSTION TEST 12 14 4 2 COMMISSIONING TEST 15 4 3 PRESSURE TEMPERATURE TESTS 16 17 4 4 LET BY AND TIGHTNESS 18 4 5 ROOM CO TEST 20 4 6 KANE4...

Страница 3: ...2 2 RETURNING YOUR ANALYSER TO KANE 42 12 3 WHERE TO SEND YOUR ANALYSER 43 13 COLD WEATHER PRECAUTIONS 43 14 ANALYSER SPECIFICATION 44 45 15 ELECTROMAGNETIC COMPATIBILITY 46 16 END OF LIFE DISPOSAL 46...

Страница 4: ...and is supplied with a flue probe with integral temperature sensor A low flow detection system warns of low flow and switches the pump off This also helps to prevent water ingress from overfilled wate...

Страница 5: ...T FEATURES Battery Compartment behind rubber cover Tasklight and infra red emitter Function buttons x 4 Rotary Switch Battery Charging indicator Water Trap Particle Filter Enter Menu controls Scroll u...

Страница 6: ...Page 6...

Страница 7: ...ective rubber cover Switch the analyser on and check that the analyser s time and date are correct To reset see USING THE MENU Section 5 Charging NiMH Batteries Ensure that you use the correct charger...

Страница 8: ...onnect the flue probe s temperature plug to the T1 socket check the plug s orientation is correct see Page 6 2 1 FRESH AIR PURGE Position the flue probe in fresh air then press The analyser s pump sta...

Страница 9: ...strument This analyser must only be used in well ventilated locations by trained and competent persons after due consideration of all the potential hazards Users of portable gas detectors are recommen...

Страница 10: ...e measured temperature from the inlet probe will be used as the inlet temperature If an inlet temperature probe is not connected to the analyser during countdown the measured temperature from the flue...

Страница 11: ...ay shows CAL ZERO Always disconnect the pressure hose before zeroing Printing Data Press and quickly release to start the analyser printing The analyser displays a series of bars until this is complet...

Страница 12: ...are stable The rotary switch can be used to display the following information RATIO Display NAT GAS Fuel type can be changed via Menu R 0 0008 CO CO2 ratio CO 52p Carbon monoxide ppm CO2 6 3 Carbon d...

Страница 13: ...r send to PC via optional Wireless module Hold for 2 seconds to log a full combustion report AUX display P 0 00 The AUX auxillary display can be customised via MENU SCREEN AUX R 0 0008 The parameters...

Страница 14: ...ne Press to select this line The side lights will flash Use or to scroll or change the Log No If only one report is logged number will not change Press to confirm a Log No The side lights will stop fl...

Страница 15: ...the RATIO CO and CO2 levels are measured These levels must be as per manufacturers instruction Where manufacturers instructions are not available the CO must be less than 350 ppm and the RATIO must be...

Страница 16: ...turn Temp T 34 9C Differential Temp Press to print a full pressure test or send to PC via optional Wireless module Hold for 2 seconds to log a pressure report Viewing printing a logged pressure temp t...

Страница 17: ...Page 17 If using larger bore tubing when performing pressure tests Push orange tube over the rim of the spigot to ensure a gas tight seal This may not produce a gas tight seal...

Страница 18: ...ically stored in the memory PR1 10 15m Pressure at start of let by test PR2 10 15m Real time pressure reading TIME 59 Let by default time is 1 minute Can be changed via Menu If the let by test fails s...

Страница 19: ...sure at end of tightness test PRINT Press to print the complete test Viewing printing a logged Let by and Tightness test Select MENU REPORT TIGHTNESS VIEW Use or to select the log number to be printed...

Страница 20: ...minute test with results stored every minute LIMIT 10ppm ALARM 30 ppm TYPE C SEALED APPLIANCE 15 minute test with results stored every minute LIMIT 10ppm ALARM 30 ppm TYPE B BOILER OPEN FLUE 15 minut...

Страница 21: ...topped earlier the Room CO test will automatically end after the designated time The CO test series is automatically stored in the memory as a log number When completed the log can be printed immediat...

Страница 22: ...select the line you want and the LEDs will start to flash Now use the keys to change the number the TEST number or the LOG number Press when you are happy with the changes The LEDs will stop flashing...

Страница 23: ...12 23 59 CAL DUE ON 18 12 16 CAL DUE ON 18 12 16 CAL DUE ON 18 12 16 COMBUSTION COMMISSION TEST PRS TEMP FUEL TYPE NAT GAS PRS mbar 18 01 ANALYSER ZERO T1 o C 75 5 CO2 9 0 CO2 0 00 T2 o C 65 2 O2 5 1...

Страница 24: ...11 TIME 12 25 27 CAL DUE ON 18 12 16 CAL DUE ON 18 12 16 CAL DUE ON 18 12 16 LET BY TEST PRS 1 mbar 10 80 ROOM CO ROOM CO PRS 2 mbar 10 78 SWEEP TEST TYPE C LET BY MINS 1 00 SEALED LIMIT 10ppm APPLIA...

Страница 25: ...UR COMPANY NAME PHONE NUMBER HERE SERIAL NO 9876543210 LOG NO 01 DATE 19 01 16 TIME 09 46 53 CAL DUE ON 18 12 16 AUX FUEL TYPE LIGHT OIL CO2 0 53 CO n ppm 02 CO CO2 0 0000 O2 20 2 FLUE o C N F INLET o...

Страница 26: ...ntered will be ignored MENU SETUP UNITS SCREEN REPORT SERVICE As you scroll up or down the side LEDs illuminate to point to the active line MAIN MENU SUB MENU OPTIONS COMMENTS SETUP Language English S...

Страница 27: ...ELLETS Fuel Origin UK N AMERICA FRANCE EFFICIENCY GROSS NET GROSS COND NET COND PRESSURE See next table below GAS ppm ppm n TEMP C F O2 REF Up down to set value 3 default NOx CALC Up down to set value...

Страница 28: ...n the side LEDs illuminate to point to the active line NOTE To EXIT the MENU at any time simply move the rotary switch to any position other than Menu Any changes that have not been entered will be ig...

Страница 29: ...ou scroll up or down the side LEDs illuminate to point to the active line NOTE To EXIT the MENU at any time simply move the rotary switch to any position other than Menu Any changes that have not been...

Страница 30: ...he flow temperature sensor T2 21 3C Use the T2 connection for the return temperature sensor T 0 0C Real time temperature difference The standard printout for this mode is as follows KANE456 SW19392 V0...

Страница 31: ...Page 31 If using larger bore tubing when performing pressure tests Push orange tube over the rim of the spigot to ensure a gas tight seal This may not produce a gas tight seal...

Страница 32: ...lar Do not exceed the flue probe s maximum temperature 600o C Do not exceed the analyser s internal temperature operating range Do not put the analyser on a hot surface Do not exceed the water trap s...

Страница 33: ...s not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperature or Efficiency calculation incorrect Ambient temperature set wron...

Страница 34: ...esh air purge on a KANE456 From first switch on if cold more than 5o C from the temperature at which calibrated 90 secs From first switch on if warm within 5o C of the temperature at which calibrated...

Страница 35: ...when it s time to Service Calibrate Recertify Our new simple online booking in procedure ensures you receive the fastest turnaround times possible Easy access to your analyser s calibration certificat...

Страница 36: ...Page 36 An out of calibration analyser may mean your combustion reports are invalid...

Страница 37: ...e it must be registered To register click If you have already registered click to sign in From your dashboard click for the analyser you wish to organise Service Recertification then click follow our...

Страница 38: ...Page 38 2 Book your service recertification order any additional spare parts you need 3 Once happy with your order just confirm pay...

Страница 39: ...Page 39 4 See your order print out the return to Kane label Don t forget to put the label on your package and send it to us...

Страница 40: ...COUNT THE REASONS WHY ON ONE HAND 1 Fixed prices ensure you know the full cost of ownership before you buy 2 Flue Gas Analysers will drift out of calibration over time only Kane can adjust them back...

Страница 41: ...red CO CO2 O2 sensors pump battery Parts not included are optionally fitted toxic cells cases probes All internal tubing water trap particle chemical filters Recertifying to national standards issuing...

Страница 42: ...tably sized box approx 45cm x 20cm x 23cm When only returning an analyser use a container as big as a shoe box pack out empty space with newspaper Before sealing your package please ensure you followe...

Страница 43: ...warm room the next morning Condensation may form which can affect the analyser s performance cause permanent damage Electrochemical sensors used in flue gas anlysers can be affected by condensation o...

Страница 44: ...ficiency Net or Gross 2 0 99 9 0 1 1 0 reading Efficiency High C 2 0 119 9 0 1 1 0 reading Excess Air 2 0 250 0 1 0 2 reading CO CO2 ratio 2 0 0 999 0 0001 5 reading Pressure differential Nominal rang...

Страница 45: ...AA cells Chargers optional 220v charger for NiMH batteries only 12v in vehicle charger for NiMH batteries only Dimensions Weight Handset Probe 0 8kg handset with protective rubber cover 200 x 45 x 90m...

Страница 46: ...terference or switch off if possible the offending equipment during your test At the time of writing this manual January 2016 Kane International Ltd are not aware of any field based situation where su...

Страница 47: ...of 0 3 A The chargers output voltage is 9 V dc at a maximum of 0 66A The charger has no user serviceable components Only a correctly specified and rated charger must be used with the KANE456 d The KAN...

Страница 48: ...mended that the analyser is given the protection of a carry case during transportation it is not required for normal operation k The KANE456 has an initial start up delay following switch on of betwee...

Страница 49: ...iquid or solid fuels Detailed instructions regarding the changing of the filter and the emptying of the water trap are given in Section 2 of this manual m WARNING When using a KANE456 to test an appli...

Страница 50: ...ured temperature from the flue probe will be used as the inlet temperature If neither probe is connected during countdown the analyser s internal ambient temperature will be used as the inlet temperat...

Страница 51: ...is recorded when each test is printed or stored Note When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be necessar...

Страница 52: ...cy EfN Nett efficiency EfCN Nett condensing efficiency PO Pump off O2 Calculated oxygen greater than 18 so calculation is disabled N F Temperature input not fitted CAL Number of days left before recal...

Страница 53: ...Page 53 ADDENDUM Instructions for KANE456 analysers fitted with optional Nitric Oxide NO sensors...

Страница 54: ...can be customised via MENU SCREEN AUX R 0 0008 The parameters displayed on lines 1 2 3 4 5 and 6 can be set by the user CO 52p They remain the AUX parameters until changed by the user CO2 6 3 NO 100p...

Страница 55: ...the display parameter for all four lines and then EXIT Rotate the dial from MENU to AUX to display all your chosen settings PRINTING and STORING The NO reading are printed and stored in the same way...

Страница 56: ...56 NITRIC OXIDE SENSOR SPECIFICATION Gas Measurement Resolution Accuracy Range Nitric Oxide NO 1ppm 2ppm 30ppm 1 5ppm 100ppm 1 5 reading 100ppm 0 to 100 ppm Overrange to 1500 ppm 1 Using dry gases at...

Страница 57: ...L7 1GF Your Details Name Job Title Company Name Company Address 1 Address 2 Town City County Postcode Country Phone Number Fax Number Mobile Number Email Address Product Details Note Proof of Purchase...

Страница 58: ...Testo Anton Telegan TPI Other Which magazines do you read Registered Gas Engineer Gas Installer P H P I P H A M News Heating Ventilating Plumbing Heating Plumbing Monthly Your feedback is important to...

Страница 59: ...Page 59...

Страница 60: ...r buying this analyser Before use please register on our website www kane co uk Scan the QR code to go directly to Register your Product on line or complete detach and return the Product Registration...

Отзывы: