CHECKING THE ACCURACY OF

FOOD TEMPERATURE MEASURING SYSTEMS US-

ING THE KM20REF REFERENCE THERMOMETER

The KM20REF thermometer has been specially confi gured and 
calibrated to provide an accurate and convenient method of checking 
the calibration of food temperature measuring systems. Each KM20REF 
is provided with a UKAS Certifi cate of Calibration from the Comark 
UKAS accredited calibration laboratory, which is traceable to NPL 
reference standards. This certifi cate defi nes the accuracy of the 
KM20REF at designated points.

To maximise the stability of readings from the KM20REF or indeed any 
other reference thermometer it is recommended that:-

a) it is only used as a reference thermometer

b) 

it is only used in ambient temperatures of +20°C ±5°C

c) 

it is protected from extreme temperature changes, 

shock and vibration

d) 

that only an alkaline battery is used and that this is 

replaced after 25 hours of use.

Food temperature measuring systems comprise two parts: the hand-
held meter and the sensor (probe). It is important that the accuracy of 
the complete system (meter and probe) is regularly checked against the 
reading from a reference thermometer. To maintain traceable records 
both the meter and probe should be given unique serial numbers.

To compare the reading from a food thermometer with the reference 
thermometer the following procedure may be applied:

Note:

  A separate procedure has been devised for National Health 

Service catering operations, copies of which are available from Comark 
Ltd - Part Number 12442.

Make sure that both the food thermometer and your KM20REF have 
both been at ambient temperature (b15°C and +25°C) for at 
least one hour.

Take a glass or plastic beaker that can safely hold at least 0.5 litres of 
fl uid.

Mix cold water and if needed ice cubes made from tap water to achieve 
a mixture of water and/or ice (crushed ice is best) at a degree or so 

Содержание KM20REF

Страница 1: ...KM20REF REFERENCE THERMOMETER...

Страница 2: ...Support department or your local distributor Repeated sharp flexing can break thermocouple probe leads To prolong lead life avoid sharp bends or kinks in the leads especially near the connector This...

Страница 3: ...one of the finest in the UK Comark UKAS certificates can offer the lowest uncertainty of 0 01 C and provide independent proof of correct calibration using equipment and procedures audited by UKAS ins...

Страница 4: ...line battery is used and that this is replaced after 25 hours of use Food temperature measuring systems comprise two parts the hand held meter and the sensor probe It is important that the accuracy of...

Страница 5: ...f the reading from the food thermometer differs from the traceable reading of the KM20REF by 0 5 C or more the food thermometer and probe should be returned to the manufacturer for recalibration Note...

Страница 6: ...KM20REF SERIAL NUMBER 1234 FOOD THERMOMETER SERIAL NUMBERS INSTRUMENT 0675 PROBE 0231 KM20REF READING 8 1 C ADJUSTMENT 0 1 C TRACEABLE TEMPERATURE 8 0 C FOOD THERMOMETER READING 8 5 C DEVIATION 0 5 C...

Страница 7: ...TEMPERATURE C FOOD THERMOMETER READING C DEVIATION C ACCEPT REJECT TESTED BY CHECKED BY Auto Switch Off The KM20REF has an auto switch off function to conserve battery life The instrument will automat...

Страница 8: ...EASUREMENT RANGE Pt100 100 C to 199 9 C RESOLUTION 0 1 C ACCURACY 0 2 C OPERATING RANGE 0 to 40 C CERTIFICATE OF CALIBRATION DESIGNATED POINTS FOR ACCURACY 18 C 0 C 5 C 8 C 20 C EMC Tested to EN 61326...

Страница 9: ...8...

Страница 10: ...Norwich Norfolk NR6 6JB United Kingdom Tel 44 844 815 6599 Fax 44 844 815 6598 Email sales comarkinstruments com Website www comarkinstruments com Comark Instruments PO Box 500 Beaverton OR 97077 USA...

Отзывы: