background image

2

CALIBRATION, CERTIFICATION AND SERVICE

Certifi cation

Comark can provide certifi cates of calibration for its whole product range, to suit 
ISO 9000 and other quality assurance procedures, food hygiene regulations, 
HACCPs and environmental regulations. Comark certifi cates are produced by 
independent quality controlled processes which compare product performance 
against agreed National Standards. For peace of mind and best practice Comark 
recommend annual certifi cation/recalibration.

Two levels of certifi cation are available for infra-red temperature and non 
temperature instruments, excluding humidity:- UKAS certifi cates via an external 
accredited laboratory and NPL traceable certifi cates from the Comark calibration 
laboratory.

Three levels of certifi cation are available for contact temperature and humidity 
products and these are detailed here:

a) UKAS Temperature Certifi cation

The Comark UKAS (United Kingdom Accreditation Service) accredited 
temperature calibration laboratory is one of the fi nest in the UK. Comark 
UKAS certifi cates can offer the lowest uncertainty of 0.01°C and 
provide independent proof of correct calibration using  equipment and 
procedures audited by UKAS inspectors. The equipment used is fully 
traceable to the National Physical Laboratory.

b) UKAS Humidity Certifi cation

In addition to the Comark temperature laboratory, the humidity laboratory  
continues the tradition of high accuracy certifi cation and a wide range 
(25% to 90%RH) with uncertainties of 2.8% of reading.   This range also 
includes Dew point measurements.

c) NPL Traceable Certifi cation

Comark NPL certifi cates are traceable to the National Physical 
Laboratory and can offer uncertainty as low as 0.3°C.

Conformance

Certifi cates of conformance can be supplied for new, serviced and recalibrated 
instruments. These confi rm that instruments are within their original manufactured 
specifi cation.

Service/Repairs

Regular servicing and any required repairs, under warranty or after, are available 
from the Comark Service Department.

For more information on all Comark certifi cation, calibration and service facilities 

please call Comark Customer Support or contact your local distributor.

0451

Содержание KM20REF

Страница 1: ...KM20REF REFERENCE THERMOMETER...

Страница 2: ...Support department or your local distributor Repeated sharp flexing can break thermocouple probe leads To prolong lead life avoid sharp bends or kinks in the leads especially near the connector This...

Страница 3: ...one of the finest in the UK Comark UKAS certificates can offer the lowest uncertainty of 0 01 C and provide independent proof of correct calibration using equipment and procedures audited by UKAS ins...

Страница 4: ...line battery is used and that this is replaced after 25 hours of use Food temperature measuring systems comprise two parts the hand held meter and the sensor probe It is important that the accuracy of...

Страница 5: ...f the reading from the food thermometer differs from the traceable reading of the KM20REF by 0 5 C or more the food thermometer and probe should be returned to the manufacturer for recalibration Note...

Страница 6: ...KM20REF SERIAL NUMBER 1234 FOOD THERMOMETER SERIAL NUMBERS INSTRUMENT 0675 PROBE 0231 KM20REF READING 8 1 C ADJUSTMENT 0 1 C TRACEABLE TEMPERATURE 8 0 C FOOD THERMOMETER READING 8 5 C DEVIATION 0 5 C...

Страница 7: ...TEMPERATURE C FOOD THERMOMETER READING C DEVIATION C ACCEPT REJECT TESTED BY CHECKED BY Auto Switch Off The KM20REF has an auto switch off function to conserve battery life The instrument will automat...

Страница 8: ...EASUREMENT RANGE Pt100 100 C to 199 9 C RESOLUTION 0 1 C ACCURACY 0 2 C OPERATING RANGE 0 to 40 C CERTIFICATE OF CALIBRATION DESIGNATED POINTS FOR ACCURACY 18 C 0 C 5 C 8 C 20 C EMC Tested to EN 61326...

Страница 9: ...8...

Страница 10: ...Norwich Norfolk NR6 6JB United Kingdom Tel 44 844 815 6599 Fax 44 844 815 6598 Email sales comarkinstruments com Website www comarkinstruments com Comark Instruments PO Box 500 Beaverton OR 97077 USA...

Отзывы: