http://www.nexbill.co.kr
item
UV sensor Adjustment
6-01-3
Related problem
Special Tools
Sensor note &
UV Test note
Medium
Work time 5 mins
Standard Value
(Unit: AD value)
Adjustment Procedure
Maintenance
All models are same maintenance mode.
1. Clean count sensors
2. Enter maintenance mode and execute VID1303.
3. The display indicates left sensor AD value. Adjust left sensor AD value at
“60” with VR9.
The display indicates Right sensor AD value by “6” digits from Left sensor
AD value. Adjust Right sensor AD value at “60” with VR10
4. To continue UV(Penetration & Reflection) & RGB adjustment, Press “6” or
“4” digits.