4
1. Main unit
1 unit
2. 3.5MHz mechanical sector probe
1 PC
3. Power adapter
1 PC
4. Internal battery 2 PCS
5. Charger
1 PC
6. Support bracket
1 PC
7. Arm-band
1 PC
8. Straps
2 PCS
9. Leather bag
1 PC
10. Shutter release
1 PC
3.2 Optional parts
1. Video recorder P93W-S
2. Mouse
3. Probe holder
4. Plastic seal box
Chapter Four Operation Condition
4.1 Power supply
Mains voltage: 100-240V
~
±10
%
Mains frequency: 50/60Hz±1Hz
4.2 Operation Environment
Ambient temperature: 10℃-40℃;
Relative humidity: 30%-75% (without condensation);
Atmospheric pressure: 700hPa-1060hPa.
4.3 Storage and Transport
Ambient temperature: -20℃-55℃;
Relative humidity: 30%-93% (without condensation);
Atmospheric pressure: 700hPa-1060hPa.
Chapter Five System Installation and Check
Warning:
1.
Do not connect the three-core power supply cord to unprotected two-core socket, or electric
shock may occur.
2.
If breakers and fuse of the mains power socket are identical to those of this system, and they
Attention:
The mains voltage is varies with different countries or regions.
Attention:
System should be avoided using in following environments:
1. Moist
2. Rain 3. Thunderstorm weather
4. No ventilation 5. Close to heat source
6. Direct sunlight
7. Dramatic temperature change
8. Poisonous gas
9. Corrosive gas
10. Strong shock
11. Strong electromagnetic field (e.g. MRI)
12. Radiation (e.g. X-ray, CT)
13. Defibrillators or short wave therapy equipment
Danger:
Do not use this equipment where flammable gas (such as anesthetic gas,
oxygen or hydrogen) or flammable liquid (such as alcohol) are present. Failure to do so
may result in explosion.
Warning:
Avoid using this equipment with high frequency operational equipment, or
danger may occur.