Kaixin MSU1 Скачать руководство пользователя страница 15

 

 

10 

Chapter Six   Functional Operation 

 

6.1 Startup and Shutdown 

In shutdown status, press key, machine starts up, power indicator lights. 
In startup status, press key, machine shuts down, power indicator goes out. Please note that when 

shut down the machine, the time of pressing key is a rather long than normal pressing key. 

 

6.2 System Functions Setting 

6.2.1 Time Setting 

1.

 

Freeze image, press direction keys

to move cursor to “Preset”; 

2.

 

Press direction keys

to enter setting interface;   

3.

 

Press direction keys

to select “Minutes, Hours, Day, Month and Year”;   

4.

 

When setting minutes, hours, day, month and year, press direction key to increase value or press 

direction key to decrease value; 

5.

 

Press key to confirm this time setting and quit setting interface. 

6.2.2 TV Mode Setting 

1.

 

Freeze image, press direction keys

to move cursor to “Preset”;   

2.

 

Press direction keys

to enter setting interface;   

3.

 

Press direction keys

to move symbol “<” to point to “TV mode”;   

4.

 

Press direction keys

to realize TV mode conversion between PAL and NTSC;   

5.

 

Press key to quit setting interface. 

6.2.3 Energy Saving Setting 

1.

 

Freeze image, press direction keys

to move cursor to “Preset”;   

2.

 

Press direction keys

to enter setting interface;   

3.

 

Press direction keys

to move symbol “<” to point to “Sleep”;   

4.

 

Press direction keys

to select energy saving time among 1~99 minutes or select “Off”;   

5.

 

Press key to quit setting interface.   

6.2.4 Characters Brightness Setting 

1.

 

Freeze image, press direction keys

to move cursor to “Preset”;   

2.

 

Press direction keys

to enter setting interface;   

3.

 

Press direction keys

to move symbol “<” to point to “Font Bright”;   

4.

 

Press direction keys

to select characters brightness among 160, 192, 224 and 255; 

5.

 

Press key to quit setting interface.   

6.2.5 Hospital Name Setting 

1.

 

Freeze image, press direction keys

to move cursor to “Preset”;   

2.

 

Press direction keys

to enter setting interface;   

3.

 

Press direction keys

to move symbol “<” to point to “Hospital”; 

4.

 

Press key, the cursor is located above “NAME”; at the same time characters input menu will be 

shown at the bottom of the screen:   

      Caps 0 1 2 3 4 5 6 7 8 9 a b c d e f g h   

Shift i j k l m n o p q r s t u v w x y z   

Press direction keys or operate mouse to move cursor to point to Caps, and then press key or 

click left mouse to achieve capital and small letter conversion;  If the cursor point to Shift, 
press key or click left mouse again to achieve the conversion between the letter and punctuation; 

Содержание MSU1

Страница 1: ...nicht übersetz werden nur als Hilfsmaterial gedackt MSU1 Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO LTD Information for Users ...

Страница 2: ...prior notice Liability and Disclaimer Statement 1 This user s manual contains warnings for predictable dangers Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual Kaixin shall not be liable for the damages and losses arising out of neglecting to follow the operation instructions herein described 2 This user s manual shall be furnished with the machine so...

Страница 3: ...thquake etc 7 Damage or malfunction caused by use of unqualified ultrasonic coupling gel 8 Damage or malfunction caused by use of probe not provided by our company 3 The company will offer repair service for instrument out of warranty but additional fees of materials and repair service will be charged 4 The customer can repair the instrument out of warranty by themselves and if necessary the compa...

Страница 4: ...light injury or property loss or affecting the use if not avoided Indicates that make sure refer to relevant contents in this manual 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Fragile AC Power Stacking limit by number DC Power Temperature limits Power supply indication Humidity limitation Up Atmospheric pressure limitation Keep dry Marking for the separate collection of e...

Страница 5: ...nnect to power 7 5 8 Probe check before and after operation 8 5 9 Main unit check before and after operation 8 5 10 System reset 8 Chapter Six Functional Operation 9 6 1 Startup and Shutdown 9 6 2 System Functions Setting 9 6 2 1 Time Setting 9 6 2 2 TV Mode Setting 9 6 2 3 Energy Saving Setting 9 6 2 4 Characters Brightness Setting 9 6 2 5 Hospital Name Setting 9 6 2 6 Key Sound Setting 10 6 2 7 ...

Страница 6: ...Open image 13 6 9 Text Input 13 6 10 Case report processing center V3 1 software optional 14 Chapter Seven General Measurement 15 7 1 Distance Measurement 15 7 2 Circumference Area Volume Measurement 15 7 3 Slope Heart rate Cycle Measurement 16 Chapter Eight Obstetric Measurement 17 8 1 Measurement and Calculation items 17 8 2 Measurement of GA and EDC 17 8 3 Obstetric report 17 8 4 Measurement it...

Страница 7: ...on 10 Adjustment of near field far field total gain and dynamic range 11 Image storage function 12 Image up down left right reverses function 13 Case information image annotation and auto time display 14 Body marks 15 Measurement of distance circumference area volume slope heart rate and cycle 16 OB software package including 17 obstetric tables about 5 kinds of animals automatic calculation of GA...

Страница 8: ... echo wave and converted to electric signal again The acoustic lens is on the probe surface Supply ultrasonic coupling gel to the acoustic lens surface when performing ultrasonic diagnosis 2 Cable To connect the probe to the probe connector 3 Probe connector To connect the probe to ultrasonic diagnostic system 2 4 Function keys description SN Function keys Real time mode function Freeze mode funct...

Страница 9: ...ing 1 Do not connect the three core power supply cord to unprotected two core socket or electric shock may occur 2 If breakers and fuse of the mains power socket are identical to those of this system and they Attention The mains voltage is varies with different countries or regions Attention System should be avoided using in following environments 1 Moist 2 Rain 3 Thunderstorm weather 4 No ventila...

Страница 10: ...c diagnostic operation 3 It is forbidden to drag and press the power and probe cables emphatically regularly inspect whether there is pull apart and bareness if there is the phenomena like this turn off power supply immediately stop using it and change it for new one 4 It is forbidden to load and unload the probe or move the instrument in galvanic to avoid danger of safety 5 Pull out of the plug f...

Страница 11: ...nnecting cable after diagnostic operation Use of defective probe may cause electric shock 3 Do not knock or bump the probe or it may be damaged and cause electric shock a Take out the attached straps install the straps on the rings of main unit as fig shown b Adjust the strap length the front strap hung on the neck the back strap fixed on the waist two straps can also adjust to similar length to m...

Страница 12: ...ttery storage removed Upper cover of battery storage as shown in figure take out the battery Warning Before connecting or using the probe make sure that the probe connecting cable and connector are in normal condition free of cracks or drop Use of defective probe may cause electric shock Attention 1 Turn off the ultrasonic system before disconnecting the probe Disconnecting the probe with system p...

Страница 13: ...er end to the power supply socket 5 6 Connect to the mouse Connect the mouse to USB interface on the right side of the main unit 5 7 Connect to power 1 Connect to the power adapter Insert the output plug of adapter into power input interface of main unit 2 Connect to the main power supply Insert the power plug jack furnished with the machine into power input socket of the power adapter the other e...

Страница 14: ...jects as specified in relevant sections 7 No barriers around the intake of equipment 8 See to it that probe has been cleaned disinfected and sterilized else dispose it as specified in relevant sections 9 Check all the ports of the machine for possible damage or blockage 10 Clean the field and environment 5 9 2 Inspection after start up Check the following items after starting the machine 1 No abno...

Страница 15: ... direction keys to move cursor to Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to point to Sleep 4 Press direction keys to select energy saving time among 1 99 minutes or select Off 5 Press key to quit setting interface 6 2 4 Characters Brightness Setting 1 Freeze image press direction keys to move cursor to Preset 2 Press direction keys to enter s...

Страница 16: ... 6 3 2 B B Mode 1 In real time mode press key to enter B B mode 2 B B image switch Press key to enter main menu and press direction keys to move the cursor to B B Mode and then press direction keys to switch image display or right click mouse to switch image display the selected image is activated and the other one is frozen 3 In real time mode press key to exit B B mode 6 3 3 4B Mode 1 In real ti...

Страница 17: ...ursor to Far in the display area Press direction key to increase far field gain and direction key to reduce far field gain so as to control the gain in far field region 6 4 5 Dynamic Range Adjustment In real time mode press key to enter main menu and press direction keys to move cursor to Dyn in the display area Press direction key to increase the value of dynamic range and direction key to decrea...

Страница 18: ...ozen status press key or middle mouse key to unfreeze the image 6 6 Puncture guide line and lithotripsy positioning line Puncture guide line In real time B mode press key to enter main menu and press direction keys to move cursor to Puncture and then press direction keys to choose line 1 press key to confirm two puncture guide lines appear on the screen press direction keys to change the angle of ...

Страница 19: ...real time mode Explanation The images stored in the U disk can only be read out in your computer 6 8 2 Open image 1 Freeze the image 2 Continuously press key twice a Read prompt appears on lower right corner of the screen 3 Press direction keys to select the image code need to be read out such as choose 003 4 Press key the image stored in frame 003 is read out 5 Press key to quit reading status an...

Страница 20: ...ic images transmit to PC and realize images mass memory 2 Support five types case report template edit and print 6 10 2 Use Instruction 1 Follow the USB Installation Instruction V3 1 of CD to install the Case Report Processing Center V3 1 software 2 Insert USB cable into USB interface of main unit and then switch on the power of main unit 3 Connect the other end of USB cable to USB Interface of co...

Страница 21: ... unfreeze clear all marks and data and quit the measurement status 7 2 Circumference Area Volume Measurement Circumference area volume measurement with ellipse method 1 In B B B mode freeze desired image the cursor is located in the Measure position of display area 2 Press key the measurement methods are showed in the lower left of the screen press direction keys to choose Ellipse press key again ...

Страница 22: ...in the lower left of the screen continue to choose Trace to measure 6 Repeat steps 3 4 to complete the multi group data measurement 7 Continuously press key twice to quit the measurement status 8 Press direction keys to choose Clear and then press key or synchronously click the left and right key of mouse to clear all marks and data 9 Press key or click the middle mouse to unfreeze clear all marks...

Страница 23: ...ment The measured value is automatically displayed at the built in mark mm on the right side of the screen G A and EDC value to be displayed in real time in the right area of the screen 6 Press key or repeatedly click left mouse to exchange the starting point and end point 7 Press key or click right mouse to finish the first measurement 8 Repeat the steps from 3 to 7 to complete the multi group da...

Страница 24: ...e sound head part of the probe See sterilization sketch map in liquid for more than 10 hours for sterilization Warning Turn off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any water roof device in the system Do not splash any water or liquor into the system whe...

Страница 25: ...3 hours and up to 6 hours Over charging or discharging will shorten the battery life the full charged battery can be used for 2 3 hours 3 Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use time reduced significantly compared with normal conditions the battery should be promptly replaced 4 The excess high or low temperature will affect th...

Страница 26: ...y into the microwave oven or pressure vessel which will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off battery such as dry battery or different capability or different model or different brand battery which will lead to the leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or turn colors or...

Страница 27: ...romagnetic interference present around the machine 3 Check power plug socket of the instrument or probe connectors They shall be properly contacted Image display is not clear 1 Adjust the total gain near field far field 2 Adjust the brightness and contrast level Control panel malfunction Restart the system by turning off the main unit power 9 3 System maintenance To ensure the system performance a...

Страница 28: ...taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shock protection type Class I internally powered equipment 2 Classified according to electric sho...

Страница 29: ...uzhou Jiangsu China Zip Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail gm01 kxele com Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Information contained in this manual is subject to change without further notice ...

Отзывы: