If the following problems occur on starting up the machine, try to make corrections following the method
in the table. If the problem remains unsolved, contact International Trade Dept of Kaixin for support.
Trouble
Correction
Power light is off and no screen
display is present when starting the
machine.
1. Check power supply.
2. Check power cable and connector.
3. Check power adapter.
Character and gray scale are
displayed, but no ultrasonic image
on the screen.
Probe is not properly connected. Turn off the power and
reconnect the probe.
Intermittent stripe, snow, or
far-field interference appears on
screen.
1. Check power supply.(spark interference present)
2. Check environment. Interfering source of around the
machine, such as electric motor, ultrasonic atomizer,
automobile, computer or other interference (Electromagnetic
interference present around the machine).
3. Check power plug/socket of the instrument or probe
connectors. They shall be properly contacted.
Image display is not clear.
1. Adjust the total gain, near field, far field.
2. Adjust the brightness and contrast level.
Control panel malfunction
Restart the system by turning off the main unit power.
9.3 System maintenance
To ensure the system performance and safety, it must be checked after using 1 year. When check the
instrument, please consult the factory-authorized distributor of Kaixin Company for service, as they need
to have professional technology engineers.
Inspect
sorts
Cleaning
Electrical safety
Mechanical safety
Image recording
Inspect
items
System inside
Peripheral
equipment
Earth resistance
Earth leakage current
Enclosure leakage
current
Patient leakage
current
Dielectric strength
Inspect the control
panel
Installed device of
peripheral equipment
Other mechanical part
Probe appearance
Images in each mode
Recorded images using
typical probe