Full Digital Ultrasonic Diagnostic Instruments
User’s Manual V1.00
- 8 -
5.1 System placement
Please carefully read through and fully understand the safety cautions before moving and placing
the system.
1. Unpack the instrument case and check the goods for its completeness according to the packing
list.
2. Place the instrument on a stable and leveled position.
3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the
instrument.
5.2 Probe bracket installation
The probe bracket should be mounted behind the main unit.
5.3 Ultrasonic probe installation
5.3.1 Ultrasonic probe connection
1. Open the probe locking knob with the cable to right; insert the probe connector into the probe
socket on the right-hand of the main unit until the metal spring plates is evenly contacted.
Tightly press the connector.
2. Rotate the probe locking knob 90 degree clockwise and lock it.
3. Check if the probe connector is locked. After turning off the system, plug the probe connector
in the probe socket then rotate the locking knob clockwise to lock it.
Warning:
Before connecting or using the probe, make sure that the probe,
connecting cable and connector are in normal condition (free of cracks or drop). Use
of defective probe may cause electric shock.
Warning:
1. The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic
system only. Select proper probe according to the relevant instructions of
ultrasonic diagnostic system.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not knock or bump the probe; or it may be damaged and may cause patient
subjected to electric shock.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric
shock.
Attention:
1. Turn off the ultrasonic system before disconnecting the ultrasonic probe.
Disconnecting the probe with system power on may damage the system or probe.
2. Before disconnecting the ultrasonic probe, place the probe on the probe bracket so
that the probe may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of
service life of probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off
power supply in short time.
Attention:
Adequate space from rear, left and right side of the machine shall be
reserved, or the machine may malfunction under excessive heat inside the enclosure.