22
7
Set-up
Section 7 - Set-up
Before use, see chapter “Section 9 - Settings”.
Gas Supply
1.
The G210 is not supplied sterile and should be cleaned before use. Make sure the gas
input and output ports at the back of the incubator are also cleaned. See “Section 12 -
Maintenance” on page 43.
2.
Install the Origio Gas Line Filter, see “Replacing the Origio Gas line Filter” on page 49.
3.
Connect the gas supply via the gas connectors (1) at the back of the unit.
4.
Turn on incubator.
5.
Make sure the incubation temperature, gas mixture etc, settings are as desired using the
Settings Menu on the touchscreen. After 30 minutes the unit will be at a constant working
temperature and air flow (see “Changing the Temperature Setpoint” on page 27).
6.
Set correct date and time before connecting to the network (LAN).
WARNING:
Use only pure CO
2
and pure N
2
gas. Use of other gases could result in
serious injury. Make sure that gas supply pressures are kept between a minimum of 0.5
bar and a maximum of 1.0 bar.
Factory Settings
The G210 is supplied with the following factory settings
Criteria
Setting
Temperature
37.0° C
Gas concentration
CO
2
: 5.0%
O
2
: 5.0 %
Gas flow*
CO
2
: 1.5 l/h*
N
2
:7 l/h*
*When G210 is stable and CO
2
and O
2
gas concentration setpoints are 5.0%
1