Safety
9
2
The G210 is intended for use in the electromagnetic environment specified below. The customer or the user of
the G210 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
Group 1
The G210 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class A – Complies
The G210 is suitable for use in all establishments other than
domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A - Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Class A - Complies
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The G210 is intended for use in the electromagnetic environment specified below. The customer or the user of
the G210 should assure that it is used in such an environment.
IMMUNITY Test
IEC 61326-1
Test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+ 4 kV contact
+ 2, 4, 8 kV air
+ 2, 4 kV contact
+ 2, 4, 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast transient/
burst
IEC 61000-4-4
+/- 1.0kV Burst Potential -
Mains
+/- 0.5kV Burst Potential -
Signals/Control
+/- 1.0kV Burst
Potential - Mains
+/- 0.5kV Burst
Potential - Signals/
Control
Surge
IEC 61000-4-5
Line-Line: 0.5kV
Combination Wave (1.2μS
x 50μS Voltage, 8μS x 20μS
Current)
Line-Earth: 0.5kV & 1.0 kV
Combination Wave (1.2μS
x 50μS Voltage, 8μS x 20μS
Current)
Line-Line: 0.5kV
Combination Wave
(1.2μS x 50μS Voltage,
8μS x 20μS Current)
Line-Earth: 0.5kV & 1.0
kV Combination Wave
(1.2μS x 50μS Voltage,
8μS x 20μS Current)
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
100% Reduction; 0.5-1
50Hz Period Cycles; 0° &
180° Sync Angle
30% Reduction; 0.5s
Duration; 25 50Hz Period
Cycles; 0° & 180° Sync
Angle
100% Reduction; 5s
Duration; 250 50Hz Period
Cycles; 0° Sync Angle
100% Reduction; 0.5-1
50Hz Period Cycles; 0°
& 180° Sync Angle
30% Reduction; 0.5s
Duration; 25 50Hz
Period Cycles; 0° &
180° Sync Angle
100% Reduction; 5s
Duration; 250 50Hz
Period Cycles; 0° Sync
Angle
Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions