Safety
8
2
CAUTION
•
Caution:
Federal law restricts this device to sale by or on the order of a physician or a
practitioner trained in its use.
•
DO NOT
use the incubator if ShockWatch or TipNTell has been triggered or if the package
is damaged.
•
Read and understand the manual completely before use. Keep the manual close to
the unit.
•
Never use or handle this unit in ways other than specified in this manual. The user’s safety
may be at risk and the unit may get damaged.
•
Never try to move the unit without consulting a person authorized by CooperSurgical.
•
Never use the unit if the alarm system of the device has issued a failure message and the
cause of the failure has not been identified.
•
Protect the power cord from being damaged or being restricted in any way. Unplug the
power cord from the wall socket or at the rear of the instrument to disconnect the mains
supply.
•
Make sure that CO
2
and N
2
gas supply pressures are maximum 1.0 bar and not below 0.5
bar.
•
Always keep the red cap on unused gas inlets at the back of the unit and the protection cap
on the sample port placed behind the Preparation Chamber.
•
Never use the unit without a genuine Origio Gas Line Filter.
•
DO NOT
expose the filter to liquids. Change filters that have been exposed to liquids.
•
DO NOT
leave lids open for more than 20 seconds.
•
DO NOT
use the unit at ambient temperatures exceeding 30°C. Ambient temperature
above 30°C will compromise the incubation process. The relative humidity must not exceed
75% (non-condensing).