Service
Replacing the Origio Gas line
Filter
Remove all samples from the chambers before
replacing the filters.
1.
To access to the filter compartment, push
the safety lock lever (1) and lift the upper
part of the incubator containing the
chambers.
CAUTION:
Do not use the unit without a genuine Origio Gas Line Filter.
2.
Fasten the connectors to the inlet and
outlet of the Origio Gas Line Filter (2).
CAUTION: Contamination Hazard
As the filter may have been used for processing and treating infectious substances, it
might be contaminated. The used Gas Line Filter should be placed in a sealed plastic bag
and labeled as biohazard material, then disposed of according to local requirements.
2
3.
Ensure the direction of flow shown on the
Origio Gas Line Filter label (3) is consistent
with the flow direction (4) shown inside the
filter compartment.
4.
Place the Origio Gas Line Filter in the filter
holder with the label facing up (5).
5.
Lower the upper part of the incubator.
Disposal of the Origio Gas line Filter
1
CAUTION:
Insert filter with the filter label facing up. If the filter is not placed correctly, it
can cause excessive gas consumption.
5
4
3