42
Calibration
4.3 Saving Calibration Constants
Factory-calibrated constants are permanently stored in a bank of the
onboard EEPROM and cannot be modified. When the device is recali
-
brated through auto-calibration, the software stores the new constants
in a user-configurable section of the EEPROM. To restore original factory
calibration settings, the software can copy the factory-calibrated con
-
stants to the user-configurable section of the EEPROM. When auto-cali
-
bration is complete, the new calibration constants can be saved to the
user-configurable banks in the EEPROM.