52
9 Analog output
9.1 Response during calibration
You can choose between “following” or “unchanged” (constant).
9.2 Response of the output signal in fault condition
Depending on the configuration, the output signal can adopt the
LOW or HIGH condition in the event of a fault.
The analog outputs are configured at the operator
level (USER) or the administrator level (ADMIN) in
CO.OUT (conductivity output) and TE.OUT
(temperature output) see Chapter 6.5 “The operator
level (USER)”, page 17.
Output signal
nominal
Output signal
HIGH
Output signal
LOW
0 — 20 mA
22.0 mA
0 mA
4 — 20 mA
22.0 mA
3.4 mA
0 — 10 V
10.7 V
0 V
2 — 10 V
10.7 V
1.4 V