48
.
- When the first calibration point has been accepted, NEXT will be
shown in the bottom line.
✱
Press
✱
Temper the medium to the working temperature.
- The uncompensated conductivity is shown in the bottom line in
the LC display.
✱
The current value can be accepted immediately by pressing
(for less than 1 second).
- The temperature coefficient that was calculated is displayed.
✱
Store the value by pressing
(for more than 2 seconds) or
discard it by pressing the EXIT key.
- The instrument now shows the present conductivity (actual value).
P
P
P
Содержание ecoTRANS Lf 03
Страница 2: ......
Страница 17: ...17 6 5 The operator level USER...
Страница 18: ...18 6 6 The administrator level ADMIN...
Страница 67: ......
Страница 68: ......