8.2.2 Logic output 2 - option (submenu BIN.2)
Parameter
D
isplay
Setting range
1
Switching function
Note
Alterations of this
parameter will af
fect:
-A
L
A
R
.2
-
H
YS.2.
When the USP con-
tact or the USP pr
e-
contact is activated,
then temperatur
e
compensation FCT
.TC
is deactivated.
FCT
.2
0
= no function
1 = MAX contact: conductivity
(make, similar to LK7)
2
2 = MIN contact: conductivity
(br
eak, similar to LK8)
2
3 = MAX contact: temperatur
e
(make, similar to LK7)
2
4 = MIN contact: temperatur
e
(br
eak, similar to LK8)
2
5 = USP contact
(switching point as per USP <645>
see Chapter 11 “The USP contact (for high-
purity water)”, page 57)
6 = USP pr
e-contact
(see Chapter 11.1 “USP pr
e-alarm”, page 58)
7 = calibration timer has run down
8 = fault output
1
The default setting is shown
bold
/
2
see Chapter 10 “Relay output
/ open-collector”, page 54