49
8 Calibrating a conductivity sensor
8.3
Calibration of the temperature coefficient of the sample
medium
✱
Make preparations, see chapter 8.2 "General information", page 47.
✱
Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.
✱
Select "TEMP.COEFF. LIN.".
The current sensor temperature appears in the display (+ flashing) (1).
✱
Enter the required working temperature and confirm your entry with the
key.
Now the source of temperature acquisition can be selected (manually, or using
the temperature input of the PSU board, or the temperature input via the
optional board). This source will be active for the duration of the calibration.
An example follows: automatic temperature acquisition using the temperature
sensor integrated into the conductivity sensor.
(1)
PGM
The working temperature must be at least 5 °C above or below the reference
temperature (25.0 °C).