35
6 Operation
6.7.4
Calibration level
Depending on which operating mode has been configured (in the Basic setting
menu), one or more of the following calibration options will be available:
- Cell constant
- Temperature coefficient
6.7.5
Calibration release
Which calibration procedure may be performed directly and which may not
can be configured here, see chapter 8.2.3 "Ways to start the calibration", page
48.
6.7.6
Delete min/max values
If required, the values can be deleted once a request for confirmation has been
acknowledged,
see chapter 6.5.2 "Min/max values of the main input", page 27 or
see chapter 6.5.3 "Min/max values of the optional inputs", page 28.
6.7.7
Delete logbook
The last five calibration processes for each input are archived in the calibration
logbook. If a "Datalogger" optional board is fitted, the date and time are also
archived.
If necessary the logbook can be deleted after a confirmation prompt.
6.7.8
Delete daily batch
If required, the counter can be deleted once a request for confirmation has
been acknowledged.
6.7.9
Delete total batch
If required, the counter can be deleted once a request for confirmation has
been acknowledged.