7 Startup
42
7.13
Setting the switching differential (hysteresis) (B.HYS)
chapter 7.10 "Setting the switching function (B.Fct)", Page 39
Setting
1. Unlock the device.
chapter 7.2 "Unlocking the device (enter the code).", Page 27
2. "Turn" until "B.HYS" appears in the bottom line.
3. "Press" (- flashing, B.HYS static)
NOTE!
The value is entered "digit-by-digit“!
7.14
Setting the switching delay (B.DLY)
Setting
The switching delay is not effective for all switching edges. Only applies for JUMO DELOS SI and JUMO
DELOS HP.
1. Unlock the device.
chapter 7.2 "Unlocking the device (enter the code).", Page 27
2. "Turn" until "B.DLY" appears in the bottom line.
3. "Press" (- flashing, B.DLY static)
NOTE!
The value is entered "digit-by-digit“!
Setting range:
0 to 100 % of the nominal measuring range
Default setting:
40 % of the nominal measuring range
Setting range:
0.00 to 99.99 s
Default setting:
0.00 s