83
8 Calibrating sensors
✱
Use the
key to apply the calibration or the
key to reject the calibra-
tion.
The device returns to measurement mode.
8.3.4
Calibration logbook
The data records for the last 5 successful calibrations are saved in the con-
nected sensor, are extracted by the JUMO AQUIS 500 RS, and can be viewed
in the calibration logbook.
Accessing the calibration logbook
The device is in measurement mode.
✱
Press the
key for more than 2 seconds.
✱
Use the
key to select the "CALIB. LOGBOOK" entry.
✱
Confirm with the
key.
The data for the last successful calibration is displayed; the example shows
the values for zero point and slope:
Use the
and
keys to access all other stored calibration data records.
NOTE
When displaying the calibration data in the calibration logbook, the sensor
does not distinguish between end-value and 2-point calibration. The last
determined calibration data record is always the top entry in the list.