35
6 Operation
6.7.8
Deleting the logbook
1
The five most recent calibration procedures are archived in the calibration log-
book.
If necessary, the logbook can be deleted following a security prompt.
6.7.9
Resetting the sensor calibrations
2
In case a calibration procedure yields impermissible values, this function
restores the factory calibration of the sensors of type 202613 and 202670. The
function does not affect the calibration logbook of the sensors.
6.8
Device info
The current configuration of all important parameters (from the basic settings
menu) is listed here.
Example
TEMPERATUR SENSOR
-> MODBUS
SENSOR TYPE
-> 202613
DEVICE ADDRESS
-> 255
SALINITY
-> 10.0 g/kg
PRESSURE
-> 1013 hPa
UNIT
-> % Sat.
SAMPLE RATE
-> 5 sec
1
Only available for freely configurable sensor.
2
Only available for sensors type 202613 and 202670.