69
8 Calibrating sensors
8.1.4
Calibration logbook
The data records for the last 10 successful calibrations are saved in the con-
nected sensor, are extracted by the JUMO AQUIS 500 RS, and can be viewed
in the calibration logbook.
Accessing the calibration logbook
The device is in measurement mode.
✱
Press the
key for more than 2 seconds.
✱
Use the
key to select the "CALIB. LOGBOOK" entry.
✱
Confirm with the
key.
The data for the last successful calibration are displayed; the example below
shows the values for zero point and slope (deviation from the factory calibra-
tion) of a 2-point calibration:
As not all of the data for the 2-point calibration fits on one page in the display,
you can use the
key to access further data from the calibration data record
for the 2-point calibration:
Use the
and
keys to access all other stored calibration data records.
NOTE
Irrespective of the unit selected in the basic settings, calibrations are always
conducted in % Sat. The last determined calibration data record is always the
top entry in the list.