
8 Calibrating sensors
66
8.1.3
Two-point calibration
Requirements
• The device is supplied with voltage,
see chapter 5 "Installation", page 17 et seqq.
• The sensor is connected
• Sensor startup on the device is completed
• Calibration has been enabled, see chapter 6.7 "Administrator level", page
• the device is in "Measurement mode".
Selecting calibration methods
✱
Initiate calibration using the
key (or via the administrator level).
✱
Select "2-POINT CALIB." and start calibration with the
key.
✱
Now remove the calibration sensor from the process.
CAUTION!
The response of the outputs during calibration depends on their configuration.
Содержание AQUIS 500 RS
Страница 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Страница 8: ...2 Description 8 Block diagram...
Страница 16: ...4 Mounting 16...
Страница 111: ...111 12 Appendix 12 4 Flow diagram of the basic setting wizard 12 4 1 Automatic configuration of a sensor...
Страница 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Страница 113: ...113 12 Appendix...
Страница 114: ...12 Appendix 114 12 5 Template for panel cut out...
Страница 115: ...115 12 Appendix...
Страница 119: ......