8 Calibrating sensors
✱
Bring the sensor to the defined 0% saturation state by submerging it in an
aqueous sodium sulfite solution (concentration < 2%).
Start zero point calibration
✱
Wait until the displayed value is stable at "0.0% saturation" or has stabi-
lized near this value. This may take some time
✱
Start calibration using the
key.
The device confirms successful calibration and displays the current measured
oxygen saturation value and the current temperature on the sensor.
✱
Now reinstall the sensor in the process.
✱
Use the
key or the
key to complete the calibration process.
The device returns to measurement mode.
CAUTION!
Do not leave the sensor in contact with the sodium sulfite solution for more
than 1 hour.
NOTE
For flawless measurement, there must be no air bubbles on the bottom of the
membrane of the submerged sensor.