6 Device settings
Parameter "rAnG": selection of the display range or the measuring range
The parameter "rAnG" can only be accessed when one of the measurands
"Conductivity", "Specific electrical resistivity", or "TDS" is selected.
✱
Use the
and
keys to select the display range:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameter "CAL": preconfiguration for the calibration of the relative cell constant
The parameter "CAL" can only be accessed when the measurand
"Conductivity" is selected.
✱
Use the
and
keys to choose whether alignment should be per-
formed with a customer-specific reference solution (Edit) or with a selection
of standard reference solutions (rEF.S):
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Menu/parameter
Setting option
Meaning
Auto
Automatic measuring
range selection
Measurand
200.0 µS/cm
0.0 to 200.0 µS/cm
Conductivity
2000 µS/cm
0 to 2000 µS/cm
20.0 mS/cm
0.00 to 20.0 mS/cm
200.0 mS/cm
0.0 to 200.0 mS/cm
2.000 kohm
0.00 to 2.000 k
Ω
× cm
Specific electri-
cal resistivity
20.00 kohm
0.00 to 20.00 k
Ω
× cm
100.0 kohm
0.0 to 100.0 k
Ω
× cm
200.0 mg/l
0.0 to 200.0 mg/l
TDS
2000 mg/l
0 to 2000 mg/l
Menu/parameter
Setting option
Meaning
Edit
Customer-specific reference solution
rEF.S
Standard reference solutions