7 Calibration
7.1
Offset/zero point correction and slope correction of the
temperature input
The parameters "OFFS" (see page 29) and "SCAL" (see page 30) can be used
to carry out a zero point correction or slope correction of the temperature
input.
Reliable references (e.g. ice water, controlled precision water baths, or similar)
must be available for this purpose.
If a correction has been carried out (deviation from the default setting), this is
indicated with the message "Corr" when the device is switched on.
The default setting of the zero point and slope values is: "oFF" = 0.0, i.e. no
correction performed.
Effect of the correction on the temperature display
7.2
Calibrating the relative time constant
In addition to direct entry, the relative cell constant can also be determined
automatically via the parameter "CELL Corr" (see page 22).
7.2.1
Calibration using customer-specific reference solution ("CAL Ed-
it" function)
This function enables the user to automatically calibrate the relative cell con-
stant with a user-specific reference solution featuring a precisely identified
conductivity.
Menu selection
✱
Start configuration.
✱
Select submenu "SEt ConF".
✱
Call up parameter "CAL".
✱
Use the
and
keys to select the "Edit" setting option.
Measuring
✱
Immerse the conductivity measuring cell in the reference solution.
✱
Press the
key for >2 s.
Correction type
Device display
Offset/zero point correction
Measured value – "OFFS"
Offset/zero point correction and
slope correction
(measured value – "OFFS") × (1+ "SCAL"/100)
The main display of the device shows the measured
actual value (e.g. 1796 µS/cm) and the subsidiary dis-
play shows "CAL." with a symbol rotating around the
last digit.