25
6 Device settings
Parameter "t.Cor": setting the temperature compensation
The parameter "t.Cor" cannot be selected when the measurand "Salinity" is
selected.
✱
Use the
and
keys to select the temperature compensation:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Parameter "t.Lin": compensation coefficient
The parameter "t.Lin" can only be selected if the setting option "Lin" has been
chosen when selecting the temperature compensation ("t.Cor").
To determine the compensation coefficient see chapter 5.7.2 "Linear tempera-
ture compensation "t.Lin"", page 18.
✱
Use the
and
keys to set the temperature coefficient:
✱
Press the
key to save the setting and move to the next parameter or
✱
press the
key to save the setting and move to the next submenu of the
setting menu or
✱
press the
key to save the setting, quit the device settings, and return to
measurement mode.
Menu/parameter
Setting option
Meaning
oFF
Do not compensate conductivity measure-
ment
nLF
Nonlinear function for natural waters
according to EN 27888 (ISO 7888), e.g.
for groundwater, surface water, or drink-
ing water
Lin
Linear temperature compensation
Menu/parameter
Setting option
Meaning
0.300 to
2.000 to
3.000
Temperature coefficient in %/K