18 Technical data
18.2.6
Measuring circuit monitoring, optional boards
Input/Sensor
Underrange/
overrange
Short circuit/
sensor break
Open circuit
Special fea-
tures
pH-value (glass elec-
trode)
Yes
Configurable
Impedance mea-
surement
a
a
With impedance monitoring and detection of deposits, the sensor alarm is tripped in case of a fault.
Monitoring via impedance measurement can also be activated.
The following points should be observed to ensure correct function:
Impedance measurements are possible only with glass-based sensors.
Sensors must be connected directly to an analysis input for pH/Redox/NH
3
on the device.
Impedance converters must not be installed in the measuring circuit.
The maximal admissible cable length between sensor and device is 10 m.
Liquid resistance has a direct impact on the measurement result. It is therefore advisable to activate the im-
pedance measurement in liquids at a minimum conductivity of approx. 100 µS/cm.
Configurable
Impedance mea-
surement
a
-
pH-value (ISFET)
Yes
No
b
b
Errors in the measuring circuit (short circuit or line break) lead to display errors (underrange or overrange
or inadmissible value).
No
b
-
Resistive conductivity
Yes
No
b
Configurable
Only with
4-wire circuit
a
Inductive conductivity
Yes
No
b
No
b
-
Universal input for con-
nection of: voltage/cur-
rent signal, RTD
temperature probe
Yes
No
b
No
b
-
Universal input for con-
nection of: resistance
potentiometer
No
b
No
b
No
b
-
Содержание 202580
Страница 6: ...Content...
Страница 28: ...12 Calibrating a pH measuring chain 28...
Страница 42: ...15 Calibrating CR conductivity sensors 42...
Страница 52: ...16 Calibrating Ci conductivity sensors 52...
Страница 60: ...17 Calibrating universal inputs 60...
Страница 79: ...79 18 Technical data...
Страница 80: ...18 Technical data 80...
Страница 81: ......