16 Calibrating Ci conductivity sensors
44
TC curve (for nonlinear temperature coefficients)
If the conductivity of a liquid whose temperature coefficient changes with tem-
perature has to be measured, this method can determine 5 temperature coeffi-
cients for 5 temperature intervals. In this way, it is possible to determine a good
approximation of the temperature coefficient curve. While the operator brings
the sample solution to the temperature values requested by the device, the de-
vice determines the temperature coefficient for each interval. This requires in-
stallation of a temperature sensor that the device can use to sense the
temperature of the sample solution.
16.2.2
Calibration default settings for Ci conductivity sensors
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the Ci calibration are described in
the following.
Open the calibration default settings:
Device Settings Menu
r
Calibration
r
Select Ci Analysis Input or Universal
Input
r
Calibration Default Settings
Conductivity
6 calibration points
5 intervals
Temperature
NOTE!
The "Calibration Default Settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v
Содержание 202580
Страница 6: ...Content...
Страница 28: ...12 Calibrating a pH measuring chain 28...
Страница 42: ...15 Calibrating CR conductivity sensors 42...
Страница 52: ...16 Calibrating Ci conductivity sensors 52...
Страница 60: ...17 Calibrating universal inputs 60...
Страница 79: ...79 18 Technical data...
Страница 80: ...18 Technical data 80...
Страница 81: ......