51
16 Calibrating Ci conductivity sensors
16.3.3
Calibrating the TC-Curve
Step
Action
1
Start the desired calibration of the TC-Curve.
Device Menu
r
Calibration
r
Analysis Input 1 to 4 (Ci) and/or
Universal Input 1 to 3
r
TC-Curve
2
Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
for Ci optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operat-
ing mode, continue with step 4
3
Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
4
Bring the temperature of the sample solution to the requested set-
points in succession. Six temperatures are requested.
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
tap button to enter
measuring range
requested
temperature val-
current actual value of
temperature
Содержание 202580
Страница 6: ...Content...
Страница 28: ...12 Calibrating a pH measuring chain 28...
Страница 42: ...15 Calibrating CR conductivity sensors 42...
Страница 52: ...16 Calibrating Ci conductivity sensors 52...
Страница 60: ...17 Calibrating universal inputs 60...
Страница 79: ...79 18 Technical data...
Страница 80: ...18 Technical data 80...
Страница 81: ......