9
2.4. Technical
specifications
LC6
Adjustable temperature range
°C
-100 ... 400
Display accuracy
%
±0.5 ±1Digit
Temperature stability
°C
>±0.03
Temperature selection
digital
via keypad
indication on DIALOG-Display (LCD)
remote control via personal computer
indication on monitor
Temperature indication
MULTI-DISPLAY (LED)
DIALOG-DISPLAY
(LCD)
Resolution
°C
0.01
Absolute Temperature Calibration
ATC
INT
°C
±9.99
ATC
EXT
°C
±9.99
Temperature control
ICC - Intelligent Cascade Control, self tuning
parameters for control and modification
Electrical connections:
Computer interface RS232 or RS485
Programmer input
0 to 10 V / 0 to 20 mA
Temperature recorder outputs
Channel 1 / 2 0 to 10 V
Channel 3
0 to 20 mA / 4 to 20 mA
Stand-by input
External alarm device
24 to 0 V DC / max. 25mA
Safety sensor „SF“
Pt100, 4-lead technique
Measurement and control sensor „INT“
Pt100, 4-lead technique
Measurement and control sensor „EXT“
Pt100, 4-lead technique
Ambient temperature
°C
5 ... 40
Overall dimensions (WxDxH)
In./cm
8 x 7 x 7 / 21 x 18 x 18
Weight
lbs./kg
7.7 / 4
Control connector for solenoid valve
230 V / max. 1.25 A
115 V / max. 1.25 A
Mains power socket for heating device
resistive load
max. 3000 W
max. 1400 W
Mains power connection ±10 %
V/ Hz
230 / 50
115 / 60
Total power consumption
W
3100
1400
All measurements have been carried out at:
rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.