Remote control
44
Control connector (35)
The connector may be used with solenoid valves.
Line voltage:
230 V
∼
/ max. 1.25 A
115
V
∼
/ max. 1.25 A
11. Remote control
11.1. Setup for remote control
SERIAL
Select the „Configuration“ submenu and select the option
„Setpoint“ to define the interface (see page 26).
The interface parameters are set by selecting the submenu
„Interface“ on the programmable controller. Normally, this is a
one-time-only adjustment. (Selecting and setting menu items,
see page 38.)
Factory settings:
RS232
BAUDRATE
4800
bauds
PARITY even
parity
HANDSHAKE
Protocol
RTS/CTS
(hardware
handshake)
Data
bits
7
Stop
bit 1
)
Like all parameters which can be entered through the keypad, interface
parameters are stored in memory even after the programmable controller is
turned off.