10. Electrical connections
Notice:
Use shielded cables only.
The shield of the connecting cable is electrically connected to the plug housing.
9
6
5
1
SERIAL
RS232/RS485 serial interface (30)
This port can be used to connect a computer with an RS232 or
RS485 cable for remote control of the programmable
controller.
Pin assignments RS232:
Pin
2
RxD
Receive
Data
Pin
3
TxD
Transmit
Data
Pin 5
0 V
Signal GND
Pin 7
RTS
Request to send
Pin 8
CTS
Clear to send
Pin 1; 4; 6, 9 Reserved - do not use!
Pin assignments: RS485
Pin 3
A
Pin 5
0 VD
Signal GND
Pin 8
B
Pins 1, 2, 4, 6, 7, 9
Reserved - do not use!
Interface correspondence: RS232:
Programmable controller
Computer
9-pole plug
9-pole socket
Pin 2 RxD
⇔
Pin 3 TxD
Pin 3 TxD
⇔
Pin 2 RxD
Pin 5 GND
⇔
Pin 5 GND
Pin 7 RTS
⇔
Pin 8 CTS
Pin 8 CTS
⇔
Pin 7 RTS
Accessories:
Order No.
Description
8 980 073
RS232 interface cable 9-pol./9-pol. , 2,5 m
8 900 110