![Julabo LC6 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/julabo/lc6/lc6_operating-manual_2030694010.webp)
Operator responsibility – Safety recommendations
10
Safety installations according to IEC 61010-2-010:
Excess temperature protection
adjustable from 0 °C ... 420 °C
Supplementary safety installations
High temperature warning function
o audible (in intervals)
Low temperature warning function
o audible (in intervals)
Supervision of the working sensor
plausibility control
Alarm indication
o audible
Environmental conditions according to EN 61 010, part 1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C (for storage and transportation)
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuation of ±10 % are permissible.
Protection class according to EN 60 529
IP31
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
Caution:
The unit is not for use in explosive environment.
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class
A (industrial and commercial range).