Menu functions
36
0 V
0 mA
20 mA
4 mA
10 °C
210 °C
197 °C
202 °C
Δ Τ =
5 K
Δ Τ =
200 K
Examples:
lowest temperature value:
10 °C
highest temperature value
210 °C
Fig. shows 200
°
C scaled to paper width
rise: 50 mV/°C
lowest temperature value:
197 °C
highest temperature value: 202 °C
Fig. shows 5
°
C scaled to paper width
rise: 2000 mV/°C
EPROG - Input
This input is necessary when the nominal value is to be
determined and set by an external programmer.
Connect the external programmer to socket (21)
REG+E-PROG of the programmable controller.
•
The programmer input of the programmable controller can
be matched to the output signal of the external
programmer.
Voltage
voltage
input
Current
current
input
Select the desired input value with the P-key
and set
with enter
.
•
„L Value“ - Setting the LOW value::
Adjust and set the lowest desired working temperature on
the programmer (e.g. 0
°
C).
Enter this same temperature on the programmable
controller by pressing the appropriate buttons on the
keypad and press enter
to set.
•
„H Value“ - Setting the HIGH value:
Adjust and set the highest desired working temperature on
the programmer (e.g. 300
°
C).
Enter this same temperature on the programmable
controller by pressing the appropriate buttons on the
keypad and press enter
to set.
•
Return to the standard display by pressing escape
.