Menu functions
34
6.8. Warning functions - LL / HL -
))
))
))
More protection for your samples in the bath!
An audible signal sounds in intervals when the actual temperature
value exceeds one of the set limits (patented).
Setting the lower limit
•
With the edit keys
change the menu function from
“LL 0” to "LL 1" and then press enter
.
•
Using the cursor keys
and the edit keys
set the lower limit (example: 15.0 °C) and then press enter
.
LL (Low Limit) factory-setting: -55.0 °C
Setting the upper limit
•
With the edit keys
change the menu function from
“HL 0” to "HL 1" and then press enter
.
•
Using the cursor keys
and the edit keys
set the upper limit (example: 250.0 °C) and then press enter
.
HL (High Limit) factory-setting: 355.0 °C
Note:
The warning functions will only be triggered when the actual bath temperature,
after start from the „OFF“, „rOFF“ or „EOFF“ mode, lies within the set limits for
3 seconds.