Operating procedures
14
5. Operating
procedures
5.1. Installation
•
The unit should be set up at a dry location.
•
Place the unit in an upright position and do not obstruct
the ventilation.
•
A wall distance of at least 10 cm must be maintained
for ventilation, allowing internal heat
to be conducted away from the unit.
•
If one or more temperature controllers are set up in a
cabinet for example, take care of good ventilation
(waste heat per unit = approx. 60 Watts).
•
The ambient temperature must not exceed 35 °C.
Ambient temperatures above 35 °C result in a failure of
the unit.
•
Do not set up the unit in the immediate vicinity of heat
sources and do not expose to sun light.
5.2. Power
connection
Caution:
•
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
•
The power supply plug serves as safe disconnecting device from the line
and must be always easily accessible.
•
Never operate equipment with damaged mains power cables.
•
Regularly check the mains power cables for material defects (e.g. for
cracks).
•
We disclaim all liability for damage caused by incorrect line voltages!
Make sure that the line voltage and frequency match the
supply voltage specified on the type plate.
Deviations of ±10 % are permissible.