Operation
54
4. Use the arrow keys to select
<On>
and confirm with
[OK]
.
The display flashes briefly.
The ATC function is activated. The stored correction values are offset
against the temperature measured internally.
8.10 Device Reset
The function resets the device to the factory settings.
1.
The device is switched on.
1. Press the
[MENU]
key.
2. Use the arrow keys to select
<conF>
and confirm with
[OK]
.
3. Use the arrow keys to select the menu option
<init>
and confirm with
[OK]
.
4. Use the arrow keys to select
<YES>
and confirm with
[OK]
.
The display flashes briefly.
The device is reset to factory settings.
i