Initial operation
43
5. Select
<On>
or
<OFF>
and confirm with
[OK]
.
<On>
when device combination is connected to same power supply. The
heating power limit is activated. Depending on the total current
consumption, it limits the heating power of the circulator and prevents
overloading of the power supply.
<OFF>
when connected to disconnected power supply. Full heating
capacity is available.
The power supply for the refrigerated circulator is set up.
7.6 Adjusting the flow of the bath fluid
The flow direction of the bath fluid is set using the lever on the front control:
• Lever position left (external): The bath fluid is mainly fed into the external
application.
• Lever position right (internal): The bath fluid is mainly circulated in the
internal bath.
NOTE
Bath fluid squirts out!
If the pump connections are not sealed, bath fluid can squirt out even with
the flow direction set to internal.
• For internal temperature control, close the external pump connections
The device is switched off.
Bath fluid is filled.
1. Set the lever to a low level for internal circulation (e.g. position 2 from the
right stop).
2. Switch on the device and check whether the flow of the bath fluid suits your
application.
3. If necessary, adjust the flow until it suits your application.
The flow of the bath fluid is set.