background image

Maintenance 

 

26 

 

 

 

 
3.  Use a vacuum cleaner to carefully vacuum the dirt off the condenser. 

  Make sure that the fins of the condenser are not damaged. 

4.  Place the venting grid back on. 

  The condenser is cleaned. 

 

 

7.6  Technical Service 

 

If the unit shows faults you cannot resolve, please contact our Technical Service. 
 

 

JULABO Technical Service 
Tel.: +1(610) 231-0250 Option 3 
Fax: +1(610) 231-260 
Email: [email protected] 
 

 

Before sending a device to Technical Service, the following points must be 
observed: 
•  Clean and decontaminate the device properly to avoid endangering service 

personnel. 

•  Include a brief description of the fault. 
•  Package the device safely for shipment. 
 

 

 

 

Содержание CORIO CP-310F

Страница 1: ...Refrigerated bath basic units Original operating manual 1 950 0170 us V07 05 2022...

Страница 2: ...nt of this operating manual is protected by copyright Information including texts images and other contents may not be reproduced distributed transmitted stored or otherwise used in any form without p...

Страница 3: ...refrigerants 9 4 5 Safety symbols 10 5 Product description 11 5 1 Function description 11 5 2 Possible combinations with circulators 11 5 3 Operating and functional elements 12 5 3 1 External inverter...

Страница 4: ...ent choice JULABO would like to thank you for the trust you have placed in our company and products This operating manual will help you become acquainted with the use of our units Read the operating m...

Страница 5: ...n the operating manual of the associated circulator 2 2 Original JULABO spare parts Hassle free continuous operation and safety also depend on the quality of the spare parts used Only original JULABO...

Страница 6: ...a danger with a low level of risk which if it not prevented may result in minor to moderate injuries WARNING This signal word designates a danger with a medium level of risk which if it not prevented...

Страница 7: ...f the unit so that the operator can operate the unit safely and avoid misuse This is a laboratory device intended for temperature control applications of liquid media It can only be operated in conjun...

Страница 8: ...and the device The operator must ensure that the device is checked for safety and functionality at regular and usage related intervals The operator must ensure that the mains supply has a low impedan...

Страница 9: ...for refrigerants The device works with a refrigerant To ensure safe handling the operator must comply with the following safety instructions Details of the refrigerant to be used can be found on the...

Страница 10: ...included with the device which should be attached to the device before initial operation Safety symbols Description Warning of a danger zone Note operating manual Warning about hot surface Read opera...

Страница 11: ...precisely cool and heat samples over a wide temperature range The device combination is suitable for external temperature applications The cooling machine is controlled by the connected circulator 5 2...

Страница 12: ...lements and their position on the unit Front without cover and rear e g 310F 1 Recessed grip 2 Bath tank drain opening 3 Drain valve 4 Drain hose for 310F and 450F 5 Condenser 6 Recessed grip 7 CAN pl...

Страница 13: ...0 V voltage variant of 310F and 450F the inverter is mounted on the rear side Rear side of 310F 450F 100 V voltage variant 1 Inverter 2 CAN plug for connection with a circulator 3 Mains input socket 4...

Страница 14: ...s designed for connection to a separate power transformer or generator Connection to a low voltage line is not permitted In accordance with IEC 61010 1 the device is designed for safe operation under...

Страница 15: ...nnection 100 V 50 60 Hz 115 V 60 Hz 230 V 50 60 Hz Current consumption max A 4 4 2 Mains fuse reset A 10 10 10 Technical data 201F Performance values Lowest temperature C 20 Refrigerants R134a Dimensi...

Страница 16: ...ction 100 V 50 60 Hz 115 V 60 Hz 200 230 V 50 60 Hz Current consumption max A 4 4 3 Mains fuse reset A 10 10 10 Technical data 450F Performance values Lowest temperature C 30 Refrigerant R449A R290 Di...

Страница 17: ...0 Hz Current consumption max A 11 7 3 4 Mains fuse reset A 12 12 10 at 100 V 50 60 Hz at 100 V Technical data 601F Performance values Lowest temperature C 35 40 Refrigerants R449A R452A Dimensions Dim...

Страница 18: ...230 V 50 60 Hz Current consumption max A 8 5 Mains fuse reset A 14 10 depending on the circulator used at 100 V Technical data 1000F Performance values Lowest temperature C 40 50 Refrigerants R449A Di...

Страница 19: ...t temperature C 38 Refrigerants R449A Dimensions Dimensions W x D x H cm 45 x 64 x 74 Usable bath opening W x D cm 35 x 41 Bath depth cm 30 Volumes min max l 42 0 56 0 Weight kg 70 7 Mains connection...

Страница 20: ...04H Bath tank drain opening 1 4301 304H Bath tank bath cover gasket FKM Viton O ring drain valve FKM Viton 5 4 2 Refrigerant For safety reasons and in case of leakage in the refrigerant cycle there is...

Страница 21: ...evice against tipping in the center of the transport trolley 4 Place loose parts for the device such as cables on the transport trolley The device is then ready for transport and can be safely transpo...

Страница 22: ...ements Insufficiently dimensioned power cords must not be used See type plate for mains voltage and current value We recommend only using original JULABO spare parts 7 2 Check safety symbols The follo...

Страница 23: ...signs order no 3 383 2630 3 Note order no 3 383 2640 1 Check the safety signs on the device for legibility and completeness 2 Replace defective or missing safety signs Safety signs can be reordered fr...

Страница 24: ...g Before draining the device let it cool to room temperature Avoid direct contact with hot bath fluid Wear protective gloves The device is tempered to room temperature and switched off 1 Place a suita...

Страница 25: ...chnical service for alternative cleaning mediums The device has now been cleaned 7 5 Cleaning condenser From time to time the condenser on the front of the device should be cleaned to maintain full co...

Страница 26: ...If the unit shows faults you cannot resolve please contact our Technical Service JULABO Technical Service Tel 1 610 231 0250 Option 3 Fax 1 610 231 260 Email Service julabo us Before sending a device...

Страница 27: ...not include damage to the product resulting from accident misuse improper installation or operation unauthorized or improper repair replacement or alteration including but not limited to repairs repl...

Страница 28: ...th above Seller shall have no further liability or obligation to Buyer therefor In no event shall Seller s liability under the Initial Warranty exceed the original purchase price of the product which...

Страница 29: ...to any other provision of the Warranty Provisions nor as to any prior or subsequent breach of the same provision Freight Seller will arrange and pay for shipping and handling for the return of the un...

Страница 30: ...ional or regional regulations The circulator combination is switched off and disconnected from the mains voltage 1 Allow the device to cool down to room temperature 2 Disconnect all power cables and i...