Preparations
62
5.3. Filling / Draining
Filling
:
•
Maximum filling level: 6 cm below the bath rim
•
Minimum filling level: 7 cm (level approx. 3 cm above
surface (16) for the placement of items/fixtures)
Note:
The working filling level depends on size and number of the
items (fixtures) to be placed inside.
Fill to minimum level only. Insert the complemented shaking
carriage and correct the filling level (adding or removing
liquid) as required
Use the water bath cover to keep temperature losses to a
minimum:
Lift-up Makrolon
cover see page 64
Draining
:
•
Switch off the shaking water bath with the mains switch
and move the equipment to the table edge.
•
Place a suitable collecting bucket or tub underneath the
equipment for draining the used liquid.
•
To drain the liquid open the dainage screw (11) on the side
of the water bath.
•
After the liquid has been fully drained, securely tighten the
drainage screw (11) again.
Warning
:
There are thermal dangers when opening the bath cover:
Burn, scald, hot steam, hot parts and surfaces that can be touched.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining
(by switching the unit on for a short moment, for example).
Содержание BAT7146
Страница 12: ...Verantwortung des Betreibers Sicherheitshinweise 10 2 2 EG Konformit t...
Страница 54: ...Operator responsibility Safety instructions 52 2 2 EC Conformity...
Страница 86: ...Cleaning repairing the unit 84...
Страница 96: ...Responsabilit de l utilisateur Informations de s curit 94 2 2 C E E Conformit...
Страница 128: ...Nettoyage de l appareil Entretien 126...
Страница 138: ...Responsabilidad del propietario aspectos relacionados con la seguridad 136 2 2 Conformidad CE...
Страница 170: ...Limpieza reparaci n del aparato 168...