Shaking Water Baths
55
E
ngl
is
h
Safety installations according to IEC 61010-2-010:
Excess temperature protection
105 °C - fixed value
Classification according to DIN 12876-1
class I
Alarm indication
o audible (continuous tone)
Supplementary safety installations
High temperature warning function
o audible (in intervals)
Low temperature warning function
o audible (in intervals)
Timer
audible (in intervals)
Environmental conditions according to EN 61 010, part 1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C (for storage and transportation)
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuation of ±10 % are permissible.
Protection class according to EN 60 529
IP 21
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
Caution:
The unit is not for use in explosive environment
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).
Содержание BAT7146
Страница 12: ...Verantwortung des Betreibers Sicherheitshinweise 10 2 2 EG Konformit t...
Страница 54: ...Operator responsibility Safety instructions 52 2 2 EC Conformity...
Страница 86: ...Cleaning repairing the unit 84...
Страница 96: ...Responsabilit de l utilisateur Informations de s curit 94 2 2 C E E Conformit...
Страница 128: ...Nettoyage de l appareil Entretien 126...
Страница 138: ...Responsabilidad del propietario aspectos relacionados con la seguridad 136 2 2 Conformidad CE...
Страница 170: ...Limpieza reparaci n del aparato 168...