Appendix A - PureView Specifications 

PureView Power Source: 

It is highly recommended that all models of PureView Pulser-Receivers be supplied with USB power from an 
independently-powered USB hub or from a PC’s USB port that offers enhanced-power.    A powered hub 
capable of supplying at least 1A and preferably 2A of current to the PureView instrument is sufficient.    This 
level of current is required for supplying power to the PureView pulser, especially as the PureView instrument 
powers up. 

Pulser 

Pulse Type

 

Negative Spike Pulse 

Pulse Initial Transition

 

Model Dependent 

Pulse Amplitude

 

Model Dependent 

Pulse Energy

 

Model Dependent 

Pulse Duration

 

Model Dependent 

Damping Values

 

Model Dependent 

Pulse Repetition Freq (PRF)

 

Model Dependent 

Pulser Trigger Source

 

Internal or External, user selectable, rising-edge triggers Pulser   

External Trigger Input Signal

 

1V to 5V positive-going pulse 

Sync-Signal Output Pulse

 

Rising edge transition <3ns, Width < 600ns, +4V minimum into a 50 Ω load 

Delay, Trigger Edge to Pulser Fire

  160 ns typical 

 

Receiver 

Gain

 

-11.5 to 70 dB in 0.5 dB steps 

Phase

 

0° (noninverting) 

Bandwidth

 

Model Dependent 

High Pass Filter

 

Model Dependent 

Low Pass Filter

 

Model Dependent 

Receiver Spot Noise   

Typically < 1.3nV/Sq.Rt.Hz, input-referred 

Input Modes

 

Pulse-Echo or Through Transmission, user selectable 

Isolation between Inputs

 

60dB typical 

Receiver Input Impedance

 

50 Ω 

Receiver Output Impedance

 

50 Ω (Receiver output should be terminated in 50 Ω) 

Linear Output Signal Range

 

+500 mV to -500 mV with 50 Ω Termination.    Signals beyond this range 
many exhibit some compression of signal peaks. 

Receiver Protection (Blanking)

 

Start: 70ns (typical) following Rising Edge of Trig / Sync signal. 

 

Duration: PureView model dependent, ranging from 300ns minimum to 
1.2us maximum. 

 

Содержание PureView Series

Страница 1: ...PureView Series USB Pulser Receiver Operator Manual PureView User Manual Version 1 3 January 2021 Copyright 2021 All Rights Reserved...

Страница 2: ...poration in the United States and or other countries Safety Advisories English The PureView system is designed for manufacturing laboratory and office use Fran ais French Le syst me PureView est con u...

Страница 3: ...dirigido la seguridad total podr a potencialmente ser comprometida Deutsch German Verwenden Sie die PureView Instrumentenausr stung wie durch den Hersteller verwiesen Wenn Sie verwenden nicht wie ver...

Страница 4: ...OPERATION 9 PUREVIEW INSTRUMENT FUNCTIONAL BLOCKS 9 Control Logic and USB Interface 10 Pulser Energy and Damping 10 Pulser Voltage 10 T R and Through Select 11 Amplifier Attenuator 11 Low Pass Filter...

Страница 5: ...PULSER CONTROLS IN JSR NET CONTROL PANEL 19 Trigger Enable 19 Trigger Source 19 PRF 19 Pulser Voltage Supply Enable 20 Pulser Voltage 20 Pulser Energy 20 Damping 20 Trigger Zin 20 INSTRUMENT STATUS I...

Страница 6: ...PureView Pulser Receiver Operator Manual Page 6 Imaginant Inc Miscellaneous 23 Environmental 23...

Страница 7: ...ber they must ship the defective product to Imaginant The RMA number must be clearly marked on the outside of the shipping package The customer is responsible for the costs associated with shipping pa...

Страница 8: ...to Mini B Additional Equipment Required Customer Supplied 1 Computer running Windows O S with an available USB port preferably an enhanced power USB port 2 Ultrasonic Transducer s 3 Trigger signal sou...

Страница 9: ...s water resistant although care should be taken to keep the instrument dry Connections are easily made to PureView hardware by attaching coaxial cables to PureView s four female SMA interface connecto...

Страница 10: ...parameters such as pulse duration pulse energy and pulse amplitude Note that a change to either the Energy or the Damping value will affect all three pulse parameters The pulser produces constant amp...

Страница 11: ...ides control over the PureView receiver s gain The receiver gain can be set in 0 5dB increments across the gain range of 11 5dB to 70db Low Pass Filters The low pass filter enables a user to reduce th...

Страница 12: ...the T R connector to excite the transducer The transducer converts energy from the electrical excitation pulse into an ultrasonic pulse that is propagated into a test material Ultrasonic return signal...

Страница 13: ...oaxial cable The PureView Receiver can provide a linear output signal of 500mV maximum into a 50 load and signals of interest should not be amplified beyond these limits Over amplification of signals...

Страница 14: ...ED indicators that illuminate to indicate the instrument status as shown in the image below The Power and Pulse LED indicators indicate the following Indicator Function Power The Blue LED indicates th...

Страница 15: ...e blue power LED indicator on PureView should illuminate The first time that a PureView unit is connected to the PC there may be a slight delay as the operating system finds and loads the hardware dri...

Страница 16: ...ing figure Outpu t Trig Sync Through T R Sig Trig Oscilloscope Transmitting Transducer Receiving Transducer Figure 3 Through Mode Operation PureView Operation The following sequence of actions describ...

Страница 17: ...pulses bracket the time when the pulser fires The first short pulse occurs approximately 70ns after the rising edge of the Trig Sync signal The second short pulse follows the first by 300ns or more de...

Страница 18: ...an example the image below shows the output signal of a PureView unit with the receiver gain set to 70dB The horizontal time scale is 500ns major division The action of the receiver protection circui...

Страница 19: ...odel the high pass filters may have values of 1MHz 5MHz 30MHz 65MHz or other values Pulser Controls in JSR Net Control Panel Trigger Enable This control allows a user to enable or disable the Pulser f...

Страница 20: ...connector to either 50 or High Z 18k Selecting a value of 50 connects an internal 50 terminating impedance to the Trig Sync connector which is appropriate for terminating a 50 coaxial cable You may se...

Страница 21: ...a return is recommended the Representative will ask for details regarding the repair issue and will issue an RMA number If possible when returning equipment to Imaginant please package the equipment...

Страница 22: ...Trigger Source Internal or External user selectable rising edge triggers Pulser External Trigger Input Signal 1V to 5V positive going pulse Sync Signal Output Pulse Rising edge transition 3ns Width 60...

Страница 23: ...ity low loss USB cable should be employed to avoid any voltage drop through the cable Physical Weight 0 152kg 0 33 lbs Enclosure Dimensions 124 5mm x 50 3mm x 25 4mm 4 9 L x 1 98 W x 1 H Length includ...

Отзывы: