PureView™ Pulser-Receiver Operator Manual 

 

 

  Page 20   

Imaginant Inc. 

Pulser Voltage Supply Enable 

This checkbox in the software enables or disables the power supply voltage that is supplied to the 
Pulser circuitry. Disabling (unchecking) this control will allow all Pulser-Receiver functions to run but 
the Pulser will not generate transducer excitation pulses.    This control can be useful for debugging 
PureView-based test systems. 

Pulser Voltage 

This displays the pulser power supply voltage.    For PureView, this is typically a fixed value that cannot 
be adjusted. 

Pulser Energy 

This control selects the pulser’s energy level. This control will be disabled (greyed out) for PureView 
units that have a fixed pulse energy value. 

Damping 

This control allows the user to select between transducer-damping values. This control will be disabled 
(greyed out) for PureView units that have a fixed (single) damping value. 

Trigger Zin 

This control is used when a PureView unit is set to External Trigger mode and the external trigger 
signal is sent to the PureView unit using a 50 Ω coaxial cable.    The ‘Trigger Zin’ control allows a user 
to set the input impedance of the Trig / Sync connector to either 50 Ω or High Z (~18kΩ).    Selecting a 
value of 50 Ω connects an internal 50 Ω terminating impedance to the Trig / Sync connector which is 
appropriate for terminating a 50 Ω coaxial cable. 

You may set the Trigger Zin control to ‘High Z’ to provide a high impedance (~18kΩ) to the Trig / Sync 
connector.    This would be appropriate for trigger signals originating from a high-impedance source. 

Multiple PureView instruments set to External Trigger mode can be triggered from the same external 
trigger signal.    To do so, their Trig / Sync connectors should be daisy-chained together so that the 
trigger signal goes to a first PureView unit, then to a second, then to a third, etc.    The ‘Trigger Zin’ 
value of each of these PureView units (except the last PureView unit) must be set to ‘High Z’.    For 
proper termination of the coaxial signal, the last PureView in the daisy chain should have its trigger 
input impedance set to 50 Ω. 

Instrument Status Indicators in JSR .Net Control Panel 

Pulsing 

The pulsing status LED indicator in the JSR .Net Control Panel software will illuminate red when the pulser is 
actively pulsing, and will be grey otherwise. 

Power Limit Status 

The Power Limit Status light will be green when the PureView pulser voltage is nominal.    At high pulser firing 
rates (PRF rates) that may cause the pulser voltage to droop, the Power Limit Status light will turn red. 

Содержание PureView Series

Страница 1: ...PureView Series USB Pulser Receiver Operator Manual PureView User Manual Version 1 3 January 2021 Copyright 2021 All Rights Reserved...

Страница 2: ...poration in the United States and or other countries Safety Advisories English The PureView system is designed for manufacturing laboratory and office use Fran ais French Le syst me PureView est con u...

Страница 3: ...dirigido la seguridad total podr a potencialmente ser comprometida Deutsch German Verwenden Sie die PureView Instrumentenausr stung wie durch den Hersteller verwiesen Wenn Sie verwenden nicht wie ver...

Страница 4: ...OPERATION 9 PUREVIEW INSTRUMENT FUNCTIONAL BLOCKS 9 Control Logic and USB Interface 10 Pulser Energy and Damping 10 Pulser Voltage 10 T R and Through Select 11 Amplifier Attenuator 11 Low Pass Filter...

Страница 5: ...PULSER CONTROLS IN JSR NET CONTROL PANEL 19 Trigger Enable 19 Trigger Source 19 PRF 19 Pulser Voltage Supply Enable 20 Pulser Voltage 20 Pulser Energy 20 Damping 20 Trigger Zin 20 INSTRUMENT STATUS I...

Страница 6: ...PureView Pulser Receiver Operator Manual Page 6 Imaginant Inc Miscellaneous 23 Environmental 23...

Страница 7: ...ber they must ship the defective product to Imaginant The RMA number must be clearly marked on the outside of the shipping package The customer is responsible for the costs associated with shipping pa...

Страница 8: ...to Mini B Additional Equipment Required Customer Supplied 1 Computer running Windows O S with an available USB port preferably an enhanced power USB port 2 Ultrasonic Transducer s 3 Trigger signal sou...

Страница 9: ...s water resistant although care should be taken to keep the instrument dry Connections are easily made to PureView hardware by attaching coaxial cables to PureView s four female SMA interface connecto...

Страница 10: ...parameters such as pulse duration pulse energy and pulse amplitude Note that a change to either the Energy or the Damping value will affect all three pulse parameters The pulser produces constant amp...

Страница 11: ...ides control over the PureView receiver s gain The receiver gain can be set in 0 5dB increments across the gain range of 11 5dB to 70db Low Pass Filters The low pass filter enables a user to reduce th...

Страница 12: ...the T R connector to excite the transducer The transducer converts energy from the electrical excitation pulse into an ultrasonic pulse that is propagated into a test material Ultrasonic return signal...

Страница 13: ...oaxial cable The PureView Receiver can provide a linear output signal of 500mV maximum into a 50 load and signals of interest should not be amplified beyond these limits Over amplification of signals...

Страница 14: ...ED indicators that illuminate to indicate the instrument status as shown in the image below The Power and Pulse LED indicators indicate the following Indicator Function Power The Blue LED indicates th...

Страница 15: ...e blue power LED indicator on PureView should illuminate The first time that a PureView unit is connected to the PC there may be a slight delay as the operating system finds and loads the hardware dri...

Страница 16: ...ing figure Outpu t Trig Sync Through T R Sig Trig Oscilloscope Transmitting Transducer Receiving Transducer Figure 3 Through Mode Operation PureView Operation The following sequence of actions describ...

Страница 17: ...pulses bracket the time when the pulser fires The first short pulse occurs approximately 70ns after the rising edge of the Trig Sync signal The second short pulse follows the first by 300ns or more de...

Страница 18: ...an example the image below shows the output signal of a PureView unit with the receiver gain set to 70dB The horizontal time scale is 500ns major division The action of the receiver protection circui...

Страница 19: ...odel the high pass filters may have values of 1MHz 5MHz 30MHz 65MHz or other values Pulser Controls in JSR Net Control Panel Trigger Enable This control allows a user to enable or disable the Pulser f...

Страница 20: ...connector to either 50 or High Z 18k Selecting a value of 50 connects an internal 50 terminating impedance to the Trig Sync connector which is appropriate for terminating a 50 coaxial cable You may se...

Страница 21: ...a return is recommended the Representative will ask for details regarding the repair issue and will issue an RMA number If possible when returning equipment to Imaginant please package the equipment...

Страница 22: ...Trigger Source Internal or External user selectable rising edge triggers Pulser External Trigger Input Signal 1V to 5V positive going pulse Sync Signal Output Pulse Rising edge transition 3ns Width 60...

Страница 23: ...ity low loss USB cable should be employed to avoid any voltage drop through the cable Physical Weight 0 152kg 0 33 lbs Enclosure Dimensions 124 5mm x 50 3mm x 25 4mm 4 9 L x 1 98 W x 1 H Length includ...

Отзывы: